TDR100
1: If Flag/Port (P91)
1: 21
Do if Flag 1 is Low
2: 0
Go to end of Program Table
;Measure the datalogger power supply voltage:
2: Batt Voltage (P10)
1: 1
Loc [ Bat_Volt ]
;Measure the internal temperature of the datalogger:
3: Internal Temperature (P17)
1: 2
Loc [ DL_Temp ]
;Set port 5 high to switch ON "Switched 12V" power from the datalogger to power the TDR100.
4: Do (P86)
1: 45
Set Port 5 High
;Measure the CS610 probe connected to Channel 3 of the level 1 SDMX50 and return La/L.
5: TDR100 Measurement (P119)
1: 0
SDM
Address
2: 0
La/L
3: 3001
MMMP Mux & Probe Selection
4: 4
Waveform
Averaging
5: 1
Vp
6: 250
Points
7: 9.5
Cable
Length
(meters)
8: 5
Window
Length
(meters)
9: .3
Probe
Length
(meters)
10: .085
Probe
Offset
(meters)
11: 3
Loc [ LaL_CH3 ]
12: 1.0
Mult
13: 0.0
Offset
;Square La/L to convert to dielectric constant:
6: Z=X*Y (P36)
1: 3
X Loc [ LaL_CH3 ]
2: 3
Y Loc [ LaL_CH3 ]
3: 4
Z Loc [ WC_CH3 ]
;Multiply the dielectric constant by 0.1 to prepare for the 3rd order polynomial
;with the appropriately scaled coefficients for Topp's calibration:
7: Z=X*F (P37)
1: 4
X Loc [ WC_CH3 ]
2: 0.1
F
3: 4
Z Loc [ WC_CH3 ]
41