Section 6. Data Error Detection
6-11
1
8 bit binary
0.155ms
2,3
16 bit binary
0.216ms
4,5
32 bit binary
0.334ms
6
ASCII decimal
1.125ms
7
8 bit ASCII hex
0.226ms
8
16 bit ASCII hex
0.355ms
9
32 bit ASCII hex
0.613ms
The above timings are preliminary.
d
With the CRC32 and CRC16-CCITT there is a extra 0.19ms for an additional
calculation. This is required as part of the CRC algorithm.
The total
extra
processing time above the SDM-SIO4’s normal delays is
calculated by the following:
Total time in µsec needed before the next P113 is started =
a
+(one from
b
)+(one from
c
)+(
d
if applany other delays required
by the format command