Campbell SDM-CVO4 Features Download Page 2

© 2001, 2013

Campbell Scientific, Inc. 

October 17, 2013

Power Considerations

The SDM-CVO4 power requirements are large compared to most 
Campbell Scientific products—especially when driving signifi-
cant loads.  Care must be taken to ensure the power supply can 
deliver this higher demand.  Alkaline batteries are not recom-
mended for long-term applications.

The SDM-CVO4 has two internal power supplies—one for chan-
nels 1 and 2 and one for channels 3 and 4.  The power supply 
for channels 3 and 4 is only turned on if the datalogger sends an 
instruction that sets the output of those channels.  If channels 3 
and 4 are not used, the power consumption is approximately  
20 mA lower than when all outputs are used.

Where supported by the datalogger, and when the application 
allows it, the SDM-CVO4 can be shut down to reduce its power 
consumption to less than 0.5 mA.  In this state, all outputs are 
switched off.

Campbell Scientific, Inc.

    |    

815 W 1800 N

    |    

Logan, UT 84321-1784

    |    

(435) 227-9000

    |    

www.campbellsci.com

USA  |  AUSTRALIA  |  BRAZIL  |  CANADA  |  CHINA  |  COSTA RICA  |  ENGLAND  |  FRANCE  |  GERMANY  |  SOUTH AFRICA  |  SPAIN

Ordering Information

Synchronous Device for Measurement

SDM-CVO4

4-Channel Current and Voltage Output Module

SDM-to-Datalogger Cable

CABLE5CBL-L

5-conductor, 24 AWG cable with drain wire and Santoprene 

jacket.  Enter cable length, in feet, after the -L.   Must choose a 

cable termination option (see below).

Cable Termination Options (choose one)

-PT

Cable terminates in stripped and tinned leads for direct 

connection to a datalogger’s terminals.

-PW

Cable terminates in connector for attachment to a 

prewired enclosure.

Specifications

Weight: 363 g (13 oz)

Dimensions 
Without Mounts: 17.8 x 10.2 x 2.3 cm (7.0 x 4.0 x 0.9 in) 
With Mounts: 23.4 x 11.2 x 2.3 cm (9.2 x 4.4 x 0.9 in)
Surge: Complies with IEC61000-4-5, test level 3 (±2 kV, 2 Ω 
coupling impedance)
Operating Temperature: -25 to +50°C 
Operating Voltage: 12 Vdc nominal (8 to 16 V)
Minimum Voltage Drop Across The Internal Current Regulating 
Circuit: 2.5 V @ 20 mA current flow
Maximum Input Voltage Relative To Channel Ground: 20 Vdc
EMC Status: Complies with EN55022-1:1998 and EN50082-1:1998

Isolation

Tested Isolation: Each channel of each unit is tested for isola-

tion resistance at 500 Vdc. Pass level > 10 MΩ
Maximum Recommended Continuous Operating Voltage: 

240 Vac rms differential between an output and the data-
logger ground

Current Drain @ 12 Vdc

b

Typical Active Current: 27 or 54 mA, depending on operating 

mode (no load on output ports)

With All Outputs Off: < 0.5 mA

Current Mode

 Range: 0 to 20,000 µA
 Resolution: 5 µA
Accuracy at +23°C: ±0.02% of set c (±5 µA)

Typical Accuracy at -25° to +50°C: ±0.1% of set c (±5 µA)
Worst Case Accuracy at -25° to +50°C:  

±0.15% of full scale range + (±5 µA)

Voltage Mode

 Range: 0 to 10,000 mV
 Resolution:  2.5 mV
 Maximum Output Current: 30 mA per channel
 Minimum Load Current: 5 µA if output < 200 mV
Accuracy at +23°C: ±0.02% of set v (±2.5 mV)

Typical Accuracy at -25° to +50°C:  

±0.13% of set v (±2.5 mV)

a

 Maximum recommended continuous operating voltage assumes all issues relating to local regulations for safe installation and operation are followed (see 

Operator’s Manual Section 4.5, Safety Issues).

To estimate the total current, add the quiescent current to the sum of all output currents multiplied by 1.5. For example, if each port is at 10 mA output, 

the total = 54 + (1.5•4•10) = 114 mA.

Reviews: