background image

Q-7.1 NET RADIOMETER

6

Inspect the silica gel monthly to ensure it is still
blue and white.  If the color changes to pink
and white, replace it with dry silica gel (this
may have to be done more frequently in wet
weather).  Wet (pink and white) silica gel can
be dried by removing it from the desiccant tube
and baking at 130°C until it returns to a blue
and white color.  Remove the desiccant tube
end cap to remove and replace the silica gel.

CAUTION:  After the desiccant tube has
been replaced, install the end plug and
vinal cap.  Be sure the holes in the vinal
cap point towards the ground.

If the Q-7.1 is installed at a remote site it may
be convenient to have a spare desiccant tube
for quick replacement (be sure to remove the
outer end caps from both tube ends before
installing).  Spare desiccant tubes are available
from Radiation Energy Balance Systems.

4.2  CONDENSATION

Condensation on net radiometer windshields
causes incorrect measurements.  This is
because water does not transmit longwave
energy.  An example of this problem can be
seen by comparing measured net radiation
values from two different net radiometers at
night, one on which dew is allowed to form and
another on which dew is prevented.  Without
dew both instruments would indicate a similar
net radiation level, e.g. -50 W m

-2

.  However,

on the instrument on which dew was allowed to
form a net radiation level of about zero would
be indicated, while the other radiometer would
maintain the reading of -50 W m

-2

.  The RV2

ventilator can be used to prevent dew from
forming on the windshields.

4.2.1  Causes of Internal Condensation

If condensation develops inside the domes
check the following:

Desiccant

Make sure that the desiccant is dry (i.e. still
blue and white).

If it has turned pink, replace it with dry
desiccant or bake it until it is dry (see section
4.1 above).  If the desiccant is pink only at the
tube end nearest the sensor head, this
indicates a leak somewhere on the sensor.

Windshields and O-rings

Check that the windshields and O-rings are in
good condition and are properly seated against
the radiometer frame (see below).

If there is still no obvious reason for the internal
condensation, check for possible leaks in the
sensor as follows:

1.  Remove the desiccant tube.

2.  Immerse the radiometer in water and blow

gently into the open end of the support
arm.  A stream of bubbles will indicate the
location of any leaks.

3.  Dry the instrument with a soft facial tissue.

If the windshields get dented, reinflate
them by blowing into the end of the support
arm.

CAUTION:  When drying the windshields,
dabbing rather than wiping them will help to
prevent scratches.

4.2.2  Possible Internal Condensation
Problems After Installation

There is a small possibility that condensation
may occur on the inside of the windshields
when the Q-7.1 is first installed in the field.
This condensation is caused by evaporation of
hydrocarbon solvents used in the
manufacturing process and may be more
apparent after several hot, sunny days if
ambient temperatures are low enough.

If you notice condensation inside the
windshields after the first few days of sunshine
following installation, remove the windshields
as described below.  Allow the solvent vapor or
condensation to escape or evaporate from the
inner air space for 15 minutes then replace the
windshields.

4.3  CLEANING

The windshields may be washed with a camel
hair brush or with a paper tissue and distilled
water.

CAUTION:  Coarse paper or cloth will
scratch the windshields and should be
avoided.

Summary of Contents for Q-7.1

Page 1: ...INSTRUCTION MANUAL Q 7 1 Net Radiometer Revision 5 96 C o p y r i g h t c 1 9 9 1 1 9 9 6 C a m p b e l l S c i e n t i f i c I n c ...

Page 2: ...er warranties expressed or implied including warranties of merchantability or fitness for a particular purpose CAMPBELL SCIENTIFIC INC is not liable for special indirect incidental or consequential damages Products may not be returned without prior authorization The following contact information is for US and International customers residing in countries served by Campbell Scientific Inc directly ...

Page 3: ...te to a vertical or horizontal pipe or rod that is less than 38 mm 1 5 inches in diameter with the two larger U bolts Attach the radiometer support arm to the mounting plate with the two smaller U bolts The radiometer support arm does not need to be level Fasten the cable to the pipe or rod with tape or plastic ties to prevent strain on the wires and damage to the instrument CAUTION Ensure that th...

Page 4: ...e 500 mV range for the 21X or CR7 The slow integration with 60 Hz rejection yields a more noise free reading For still air net radiation Q can be computed from the thermopile voltage Vt by If Vt 0 Q Wm 2 Vt mV x Fp If Vt 0 Q Wm 2 Vt mV x Fn where Fp and Fn are the positive and negative calibration factors respectively The calibration factors and serial number are given on a label under the sensor ...

Page 5: ...s using fixed correction factors of 1 045 for positive fluxes and 1 000 for negative fluxes If the wind speed is measured the functions above can be programmed into the datalogger to give a real time correction of net radiation for the effects of wind speed Note that the wind speed is expressed in units of meters per second If the wind speed is measured in units other than meters per second conver...

Page 6: ...nd input location assignments On the 21X and CR7 use 500 mV input range For a Single Ended measurement use Instruction 1 3 3 PROGRAMMING EXAMPLE WITH DYNAMIC WIND SPEED CORRECTION The following CR10 X program measures an R M Young wind sentry anemometer model 03101 and a Q 7 1 The wind speed is used to give a real time wind correction to net radiation The wind speed must be measured before the win...

Page 7: ...rections 10 Beginning of Subroutine P85 1 2 Subroutine 2 11 Z X F P37 1 1 X Loc ws_m_s 2 00174 F 3 4 Z Loc A 12 Z X F P34 1 4 X Loc A 2 99755 F 3 7 Z Loc Corr_Fact Enter the negative multiplier n nnn 13 Z X F P37 1 2 X Loc Rn_mV 2 n nnn F 3 3 Z Loc Rn_W_m2 14 Z X Y P36 1 3 X Loc Rn_W_m2 2 7 Y Loc Corr_Fact 3 3 Z Loc Rn_W_m2 15 End P95 Proper entries will vary with program and datalogger channel an...

Page 8: ...ng Desiccant Make sure that the desiccant is dry i e still blue and white If it has turned pink replace it with dry desiccant or bake it until it is dry see section 4 1 above If the desiccant is pink only at the tube end nearest the sensor head this indicates a leak somewhere on the sensor Windshields and O rings Check that the windshields and O rings are in good condition and are properly seated ...

Page 9: ... lower windshields are completely seated on the frame around the entire periphery CAUTION Do not attempt to dismantle the instrument body If internal attention is needed contact Campbell Scientific Inc for repair If the windshields become dented remove the dents as follows 1 Remove the vinal cap from the support arm 2 Blow gently into the end plug breather hole to inflate the windshields Remember ...

Page 10: ...f the standard has also been verified using black body chambers Care should however be taken when comparing readings with sensors from different manufacturers or even older models from Radiation Energy Balance Systems This is because in recent years academic studies have reported inadequacies both in the design and calibration of net radiometers from most manufacturers Radiation Energy Balance Sys...

Page 11: ...This is a blank page ...

Page 12: ... campbellsci com br suporte campbellsci com br Campbell Scientific Canada Corp CSC 11564 149th Street NW Edmonton Alberta T5M 1W7 CANADA www campbellsci ca dataloggers campbellsci ca Campbell Scientific Ltd CSL Campbell Park 80 Hathern Road Shepshed Loughborough LE12 9GX UNITED KINGDOM www campbellsci co uk sales campbellsci co uk Campbell Scientific Ltd France Miniparc du Verger Bat H 1 rue de Te...

Reviews: