PVS5120-Series Portable Samplers
45
FIGURE 8-14. MD485 CS I/O tab
6. Click the
RS-485
tab. Set Baud
Rate
to
38.4k
. Set
Parity
to
None
. Set
Character Length
to
8 bits
. Set
Stop Bits
to
1 bit
FIGURE 8-15. MD485 RS-485 tab
7. Click
Apply
.
8.2.5.2 Configuring Sampler for Modbus® Control
1. Click the
Sampler Settings
tab (FIGURE
a. Set the
Control Type
to
Modbus
.
b. Set the
Modbus Address
. This is the address of the sampler and is
referenced in the
CRBasic
program running on the datalogger. The
default Modbus® address for the samplers is 1. Possible values are
1–127.
c. Set the
Number of Sampler Bottles
.
d. Set the
Samples Per Bottle
. It can range from
1
to
999
. This setting
is ignored if
Never Full
is set to
On/Enable
.
e. Typically leave the
Never Full Operation
disabled.