PVS5120-Series Portable Samplers
38
FIGURE 8-9. MD485 CS I/O tab
6. Click the
RS-485
tab. Set
Baud Rate
to 38.4 kbps. Set
Parity
to
None
.
Set
Character Length
to
8 bits
. Set
Stop Bits
to
1 bit
).
FIGURE 8-10. MD485 RS-485 tab
7. Click
Apply
.
8.2.4.2 Configuring Sampler for PakBus® Control
1. Once connected to the VSC100 in
DevConfig
, click the
Sampler Settings
tab (FIGURE
a. Set the
Control Type
to
PakBus
.
b. Set the
PakBus Address
. This is the address of the sampler and is
referenced in the
CRBasic
program running on the datalogger. The
default PakBus® address for the samplers is 95. Each sampler needs
to be uniquely addressed, as well as each datalogger on the network.
c. Set the
Number of Sampler Bottles
.
d. Set the
Samples Per Bottle
. It can range from
1
to
999
. This setting
is ignored if
Never Full
is set to
On/Enable
.