OBS501 Smart Turbidity Meter with ClearSensor
®
Technology
13
7.3.1 SDI-12 Wiring
TABLE 7-3. SDI-12 Wiring
Color
Description
CR800
CR5000
CR3000
CR1000
CR200X
Series
CR6
Red
Power
12V
12V
Black
Power
Ground
G
G
G
White
SDI-12
Signal
Control
Port
1
C1/
SDI-12
Control
Port
1
or
Universal
Channel
1
Brown
not used
not used
not used
not used
Blue
not used
not used
not used
not used
Green
not used
not used
not used
not used
Clear
Shield
G
G
G
1
Only odd control ports or universal channels can be used for SDI-12 (for example,
C1, C3...)
7.3.2 RS-232 Wiring
The CR800 series, CR1000, and CR3000 dataloggers support serial
communications with dedicated UART hardware on their control ports. They
use two control ports configured as a single communications (COMn) port. The
CR6 uses two control ports or two universal channels configured as a single
communication port.
TABLE 7-4. RS-232 Wiring
Color
Description
CRBasic
Datalogger
RS-232 9-pin
Red
Power
12V
Black
Power Ground
G
White
RS-232
Transmit
(OBS501 Tx
Output)
Rx (COM
n
, C
n
,
or U
n)
Pin 2 Rx (Input)
Blue
RS-232 Receive
(OBS501 Rx
Output)
Tx (COM
n
, C
n
,
or U
n)
Pin 3 Tx
(Output)
Brown
not used
Green
not used
Clear
Shield
G