OBS500 Smart Turbidity Meter with ClearSensor
®
Technology
14
After installing
DevConfig
, select the OBS500 in the Device Type selection.
Select the correct PC Serial Port and then click Connect (see FIGURE
).
The Terminal tab can be used to verify the setup of the OBS500. Select the
Terminal tab. Click in the Terminal window and push the Enter key several
times. This will wake up the RS-232 mode of the sensor. Once successfully
connected, you will see an OBS-500> prompt. FIGURE
DevConfig
after pressing ‘l’ (one) to identify the OBS500. By default, the OBS500 is in
the SDI-12 mode for communication. Once in the RS-232 mode, if there is no
communication for 20 seconds, the sensor will return to the SDI-12 mode.
FIGURE 7-2. Terminal Mode using 1 and H commands