Campbell HMP45C212 Instruction Manual Download Page 8

4

 

 

;Pause 150 mSec, before making measurements, so the 
;probe can stabilize on true readings. 

03:  Excitation with Delay (P22) 
 1:  1  Ex 

Channel 

 

2: 

Delay W/Ex (units = 0.01 sec) 

 

3: 

15 

Delay After Ex (units = 0.01 sec) 

 4:  0  mV 

Excitation 

 
;Measure the HMP45C212 relative humidity. 

04:  Volt (SE) (P1) 
 1:  1  Reps 
 

2: 

2500 mV Slow Range 

; CR500 (2500 mV); CR23X (1000 mV); 21X, CR7 (5000 mV) 

 

3: 

SE Channel 

; Green wire (SE 4) 

 

4: 

Loc [ RH_pct   ] 

 

5: 

.1 

Mult 

; mult 0.1 for percent RH, mult 0.001 for fraction 

 6:  0  Offset 
 
;Turn the HMP45C212 off. 

05: Do 

(P86) 

 

1: 

51 

Set Port 1 Low 

; Yellow wire (C1) 

 
 

7. MAINTENANCE

 

 

The HMP45C212 Probe requires minimal maintenance.  
Check monthly to make sure the radiation shield is free 
from debris.  The black screen on the sensor’s end 
should also be checked. 
 
When installed in close proximity to the ocean or other 
bodies of salt water (e.g., Great Salt Lake), a coating of 
salt (mostly NaCl) may build up on the radiation shield, 
sensor, filter and even the chip.  NaCl has an affinity for 
water.  The humidity over a saturated NaCl solution is 
75%.  A buildup of salt on the filter or chip will delay or 
destroy the response to atmospheric humidity. 
 
The filter can be rinsed gently in distilled water.  If 
necessary, the chip can be removed and rinsed as well.  
Do not scratch the chip while cleaning. 
 
Long term exposure of the HUMICAP® relative humidity 
sensor to certain chemicals and gases may affect the 
characteristics of the sensor and shorten its life.  Table 2 
lists the maximum ambient concentrations, of some 
chemicals, that the HUMICAP® can be exposed to.  
Detailed information on allowed concentrations can be 
requested from Campbell Scientific representatives. 
 

Table 2.  Chemical Tolerances of HMP45C212

 

 

Chemical 

Concentration (PPM) 

Organic solvents 

1,000 to 10,000 

Aggressive Chemicals 

(e.g. SO

2

, H

2

SO

4

H

2

S, HCL, CL

2

, etc.) 

 

1 to 10 

Weak Acids 

100 to 1,000 

Bases 

10,000 to 100,000 

 

Recalibrate the HMP45C212 annually.  Obtain an 
RMA number before returning the HMP45C212 to 
Campbell Scientific for recalibration. 

 

8. REFERENCES

 

 

Goff, J.A. and S. Gratch, 1946:  Low-pressure 

properties of water from -160° to 212°F, 

Trans. 

Amer. Soc. Heat. Vent. Eng.

51

, 126-164. 

 
Lowe, P.R., 1977:  An approximating polynomial 

for the computation of saturation vapour 
pressure, 

J. Appl. Meteor.

16

, 100-103. 

 
Weiss, A., 1977:  Algorithms for the calculation of 

moist air properties on a hand calculator, 

Amer. 

Soc. Ag. Eng.

20

, 1133-1136. 

 

Reviews: