Table 8-1: CS650 wiring code for RS-232 and A200
Color
Function
A200 terminal
Orange
RxD
Rx
Green
TxD
Tx
Red
Power
+12 VDC
Black
Reference
G
Clear
Shield
G
8.1.1.3 Powering the sensor
The A200 provides power to the sensor when it is connected to a computer USB port. An internal
dc/dc converter boosts the 5 VDC supply from the USB connection to a 12 VDC output that is
required to power the sensor.
8.1.1.4 Determining which COM port the A200 has been assigned
When the A200 driver is installed, the A200 is assigned a COM port number. This COM port
number is needed when using
Device Configuration Utility
. Often, the assigned COM port is the
next port number that is free. However, if other devices were previously installed (some of which
may not be plugged in), the A200 may be assigned a higher COM port number.
To check which COM port has been assigned to the A200, monitor the appearance of a new
COM port in the list of COM ports offered in your software package such as
LoggerNet
before
and after the installation, or look in the Windows Device Manager list under the ports section
(access via the control panel).
8.1.2
Device Configuration Utility
The
Device Configuration Utility
is available as a download on
.
It is included in installations of
LoggerNet
,
RTDAQ
, or
PC400
.
Connect the CS650 to the A200 as shown in
(p. 14). Connect the computer to the A200
USB port with the supplied USB cable.
Launch
Device Configuration Utility
and search for CS650 Series from the Device Type list on the
left. Select 9600 from the Baud Rate list.
CS650 and CS655 Water Content Reflectometers
14