![Campbell CS615 Instruction Manual Download Page 13](http://html1.mh-extra.com/html/campbell/cs615/cs615_instruction-manual_489212013.webp)
CS615 WATER CONTENT REFLECTOMETER
10
; loop index multiplied by 3 so that 3 readings
measured by pulse
; instruction are advanced 3 locations each pass
through measurement loop
6: Step Loop Index (P90)
1:
3
Step
7: Period Average (SE) (P27)
1:
3
Reps
2:
4
2 V Peak to Peak/200 kHz
Max. Freq.
3:
1
SE Channel
4:
10
No. of Cycles
5:
5
Timeout (units = 0.01 seconds)
6:
1--
Loc [ msec#1 ]
7:
.001
Mult
8:
0.0
Offset
8: End (P95)
; set ports enabling mux and probes low
9: Do (P86)
1:
51
Set Port 1 Low
10: Do (P86)
1:
53
Set Port 3 Low
; apply calibration
11: Polynomial (P55)
1:
48
Reps
2:
1
X Loc [ Period#1 ]
3:
49
F(X) Loc [ WatCont#1 ]
4:
-0.187
C0
5:
0.037
C1
6:
0.335
C2
7:
0.0
C3
8:
0.0
C4
9:
0.0
C5
12: Do (P86)
1:
10
Set Output Flag High
13: Real Time (P77)
1:
10
Hour/Minute
14: Sample (P70)
1:
48
Reps
2:
49
Loc [ WatCont#1 ]
15: End (P95)
End Program