Campbell CS525 Instruction Manual Download Page 10

CS525 ISFET pH Probe 

 

TABLE 3-2.  CR200(X) Series Wiring 

Wire Color

 

Wire Label/Function

 

VDIV2:1* 

 

CR200(X) Series 

 

Red  

Power 5 Vdc 

 

EX 

Black  

Signal Reference  

 

White  

Signal 

SE 

Clear  

Shield  

 

G  

 
*When using the VDIV2:1 with the CR200(X)-series datalogger, the L peg of 
the VDIV2:1 is not used and can be removed. 

4.  Programming 

4.1  Using SCWIN for CR800, CR850, CR1000, and CR3000 

In SCWIN, the CS525 is programmed by selecting single-ended voltage under 
Generic Measurements.  SCWIN is the easiest and typically the preferred 
method for programming the datalogger. SCWIN generates a wiring diagram 
that shows how to connect the CS525 to your Campbell Scientific datalogger. 

The sections that immediately follow are for CRBasic users. 
SCWIN users can jump ahead to the Calibration section. 

NOTE 

4.2  Using CRBasic 

4.2.1  Simple Program 

4.2.1.1  CR200(X) Programming 

The CS525 does not require constant power.  To conserve power, the 
datalogger can be programmed to power the CS525 prior to the measurement, 
shut off.  This function can be accomplished using the ExDelSE instruction of 
a CR200(X) datalogger.  A multiplier of 1.0 and an offset of 0.0 yields 
millivolts. A corrected multiplier and an offset are required to provide an 
output in pH units (see Appendix A). 

The ExDelSE instruction has the following format: 

ExDelSE(Dest,Reps,SEChan,ExChan,ExmV,Delay,Mult,Offset ) 

This example program uses the ExDelSE instruction to apply a voltage to 
excitation channel one, delay for 30 milliseconds, and then make a 
measurement on single-ended analog channel one. 

Summary of Contents for CS525

Page 1: ...CS525 ISFET pH Probe Revision 10 11 C o p y r i g h t 2 0 0 9 2 0 1 1 C a m p b e l l S c i e n t i f i c I n c ...

Page 2: ......

Page 3: ...alling and shipping defective products to Campbell Campbell will return such products by surface carrier prepaid within the continental United States of America To all other locations Campbell will return such products best way CIP Port of Entry INCOTERM 2010 prepaid This warranty shall not apply to any Campbell products which have been subjected to modification misuse neglect improper service acc...

Page 4: ...ng container Campbell Scientific s shipping address is CAMPBELL SCIENTIFIC INC RMA _____ 815 West 1800 North Logan Utah 84321 1784 For all returns the customer must fill out a Statement of Product Cleanliness and Decontamination form and comply with the requirements specified in it The form is available from our web site at www campbellsci com repair A completed form must be either emailed to repa...

Page 5: ... 3 1 Preparation for Use 3 3 2 Wiring 3 4 Programming 4 4 1 Using SCWIN for CR800 CR850 CR1000 and CR3000 4 4 2 Using CRBasic 4 4 2 1 Simple Program 4 4 2 1 1 CR200 X Programming 4 4 2 1 2 CR800 CR850 CR1000 and CR3000 Programming 5 4 2 2 Programming for calibration 6 5 Calibration 6 6 Maintenance 6 6 1 Cleaning 7 7 Troubleshooting 8 Appendix A Calibration A 1 Figure 6 1 Tip of the CS525 Probe 7 T...

Page 6: ......

Page 7: ...e high solids more aggressive chemicals or biological materials are present and clogging or junction contamination may occur The probe is easily cleaned using a simple toothbrush The CS525 electronics are safely embedded in the sensor body which is made of durable PEEK material Due to the elimination of the glass bulb the possibility of broken glass is eliminated The ISFET design considerably redu...

Page 8: ...tation to ensure that the expected model number and cable length were received Remember that although the CS525 is designed to be a rugged and reliable device for field use it is also a highly precise scientific instrument and should be handled as such There are no user serviceable parts and any attempt to disassemble the device will void the warranty 2 Specifications pH Range 0 to 14 Power Requir...

Page 9: ... that best fits the measurement range of the data acquisition device used Campbell Scientific s CR800 CR850 CR1000 and CR3000 can measure several output ranges With these dataloggers either the 0 to 5000 mV output or 400 mV output can be chosen for the CS525 Connect the CS525 to the CR800 CR850 CR1000 and CR3000 based on the probe s output option see Table 3 1 TABLE 3 1 CR800 CR850 CR1000 and CR30...

Page 10: ...ctions that immediately follow are for CRBasic users SCWIN users can jump ahead to the Calibration section NOTE 4 2 Using CRBasic 4 2 1 Simple Program 4 2 1 1 CR200 X Programming The CS525 does not require constant power To conserve power the datalogger can be programmed to power the CS525 prior to the measurement shut off This function can be accomplished using the ExDelSE instruction of a CR200 ...

Page 11: ...A The VoltSe instruction has the following form VoltSe Dest Reps Range SEChan MeasOff SettlingTime Integ Mult Offset Use mV5000 for the range parameter if the sensor s output option is 0 to 5000 mV Typically use mV250 for the range parameter if the sensor s output option is 300 mV The mV250 range option may be inadequate for pH ranges that are greater than 11 or less than 2 At these extreme ranges...

Page 12: ...iodically calibrate the pH reading simple program instructions can be used to make this process simple See Appendix A for an example program 5 Calibration Calibration should be carried out according to the detailed procedure outlined in Appendix A The calibration should use two or more pH standards pH4 CSI PN 25587 pH7 CSI PN 25586 and pH10 CSI PN 25588 The frequency of calibration will depend on ...

Page 13: ...r The best cleaning method is the following soapy water method 1 Place the tip of the probe in soapy water that is 50 to 60 C and soak for about 5 minutes see Figure 6 1 2 Rinse thoroughly with deionized water 3 If necessary lightly scrub the tip of the probe with a toothbrush to remove debris see Caution 4 Place the probe in saturated KC1 solution CSI PN 16349 at about 20 C Keep probe in solution...

Page 14: ...phragm is polluted Chip is scratched Suggestion Put probe in fluid Clean probe using soapy water method Probe can not be fixed if chip is scratched Problem Probe response is very slow Possible reason Diaphragm is chipped or polluted Suggestion Clean probe with soapy water method Problem Probe signal is drifting Possible reason Diaphragm can be dried out Chip is scratched Suggestion Soak probe for ...

Page 15: ... from the pH7 buffer solution and rinse with deionized water 5 Blot dry the CS525 with a soft cloth or paper towel 6 Place the CS525 in a pH 4 buffer solution CSI PN 25587 7 Allow the pHmV reading to stabilize 8 Change the value in pH4record to 1 9 Change the value in pHcal to 1 10 The CS525 is now ready to be place in the solution to be measured The simple example program to provide this two poin...

Page 16: ...1 mV5000 1 False 0 _60Hz 1 0 If pH4record 1 Then pH4mV pHmV pH4record 0 EndIf If pH7record 1 Then pH7mV pHmV pH7record 0 EndIf If pHcal 1 Then pHmult 3 pH7mV pH4mV pHoffset 7 pHmult pH7mV pHcal 0 EndIf pH pHmult pHmV pHoffset CallTable TenMin NextScan EndProg ...

Page 17: ......

Page 18: ...entific Canada Corp CSC 11564 149th Street NW Edmonton Alberta T5M 1W7 CANADA www campbellsci ca dataloggers campbellsci ca Campbell Scientific Centro Caribe S A CSCC 300 N Cementerio Edificio Breller Santo Domingo Heredia 40305 COSTA RICA www campbellsci cc info campbellsci cc Campbell Scientific Ltd CSL Campbell Park 80 Hathern Road Shepshed Loughborough LE12 9GX UNITED KINGDOM www campbellsci c...

Reviews: