![Campbell CS475 Instruction Manual Download Page 37](http://html1.mh-extra.com/html/campbell/cs475/cs475_instruction-manual_3829368037.webp)
Appendix C. SDI-12 Commands/ Changing Settings
C-3
Change the sensor’s address by sending the sensor the
aAb!
command, where
“a” is the original address and “b” is the new address. TABLE
shows an
example of the command and response for setting the address.
TABLE C-3. Example of Setting Address
Initial Command
Response
0A1!
Where 0 is the original address
and 1 is the new address.
1<cr><If>
The new address (1) is set in response.
C.1.2 Set Water Stage
With the
Set Water Stage
command, you enter the initial depth of the water,
and the sensor will automatically measure the distance between the sensor and
the water surface. The water stage setting and the initial distance measurement
are used to calculate subsequent water stage measurements. Correct stage
measurements require that the water stage setting be in the same units as the
Units of Measure setting. You can find out if the sensor is set to meters or feet
by using the
aXGU!
(
Get Unit
) command.
To set the water stage, do an
aXSS+nnn.nnn!
command (where nnn.nnn = the
initial water depth) followed by the
aD0!
(
Send Data
) command. TABLE
shows an example of the command and response for entering this setting.
TABLE C-4. Example for Setting Water Stage
Initial Command
Response
0XSS+7.010!
Where (from left to right),
0—sensor’s address;
7.010—the initial water depth value
used to calculate subsequent stage
measurements.
00011<cr><If>
Where (from left to right),
0—sensor’s address;
001—the amount of time (in
seconds) that you must wait before
sending the send data command;
1—the number of values that will be
placed in the buffer.
Subsequent Command
Command Response
0D0!
Where the first zero is the sensor
address.
This is the send data command.
0+7.010<cr><If>
Where (from left to right),
0—sensor’s address;
7.010—the initial water depth value
used to calculate subsequent stage
measurements.
C.1.3 Start False Echo Learn
The
Start False Echo Learn
command is an essential function during the
startup of the sensor. It is also used during testing if a problem is encountered.
The command allows the unit to learn about false echoes (noise) in the area.
With the
False Echo Learn
command, you enter the actual distance to the
water surface, as measured in meters or feet, depending on the unit setting. The
Summary of Contents for CS475
Page 2: ......
Page 6: ......
Page 10: ...Table of Contents iv...
Page 32: ......
Page 34: ......
Page 46: ...Appendix C SDI 12 Commands Changing Settings C 12...
Page 48: ......
Page 49: ......