CS120A Visibility Sensor
22
12. Definition of the variables that can be set by the
user on the CS120A
Both DevConfig and the command line interface can access all the user
configurable variables within the CS120A. The acceptable range and the
identification number for these variables are listed below along with a short
description.
ID
Name
Range
Description
Factory
default
1
Sensor ID
0-9
Separate ID used as an extra identifier for a
particular CS120A on a network.
0
2
User Alarm 1
Enabled
0-1
User alarm one activation state
0=Alarm one disabled
1=Alarm one enabled
0
3
User Alarm 1 Active
0-1
0=Check if distance is
less than
‘User alarm 1
Distance’
1=Check if distance is
greater than
‘User alarm
1 Distance’
0
4
User alarm 1
Distance
0-60000
Distance value that alarm one will trigger
against. This value will correspond to metres
or feet depending upon which is selected in
‘Visibility Unit’
10000
5
User Alarm 2
Enabled
0-1
User alarm two activation state
0=Alarm two disabled
1=Alarm two enabled
0
6
User Alarm 2 Active
0-1
0=Check if distance is
less than
‘User alarm 2
Distance’
1=Check if distance is
greater than
‘User alarm
2 Distance’
0
7
User alarm 2
Distance
0-60000
Distance value that alarm one will trigger
against. This value will correspond to metres
or feet depending upon which is selected in
‘Visibility Unit’
10000
8
Baud rate
0-6
Baud rate for the main RS232/RS485 interface
0 = 115200 bps
1 = 57600 bps
2 = 38400 bps
3 = 19200 bps
4 = 9600 bps
5 = 2400 bps
6 = 1200 bps
2
9
Serial Number
0-32000
Internal serial number for the CS120A.
(
Read only)
-
10 Visibility Unit
M or F
Unit the visibility value will be presented as.
M = Metres
F = feet
M
11 Message Interval
1-3600
Interval in seconds between outputs in
continuous mode. This value has no effect if
polled mode has been selected in
‘Measurement mode’
60