CRVW3 3-Channel Vibrating-Wire Datalogger
29
Use the
Save Last Results
button to save the results of the test to a binary file
(*.bin). This file can be sent to Campbell Scientific support personnel (or to
others) and then opened within
DevConfig
using the
Offline Analysis
tab that
is shown when not connected to the CRVW3. This will allow others to assess
the condition of your sensor measurement in detail. Use the
Export
button to
save the results of the test into a text file that is readable using standard text
editor programs.
The
Send OS
tab is used when updated firmware (an operating system) is
installed into the CRVW3. Campbell Scientific will periodically provide
firmware updates for the CRVW3 on its website. (see
). Once a newer firmware file is obtained
from the website, follow the installation instructions that are displayed when
this tab is selected.
You can use the
Terminal
tab to make a terminal mode connection to the
CRVW3. This is typically used under the direction of Campbell Scientific
support personnel or by advanced users. See Appendix
(p. B-1)
, for more information.
8.1.7 Validating Radio Connectivity in the Lab
For CRVW3 devices with the –RF451 or –RF407 communications option, you
should verify while still in the lab that the radio is working and has been
configured to be reasonably ready for field operation.
8.1.7.1 Validating CRVW3-RF451 Radio Connectivity
When using CRVW3-RF451 devices, the simplest way to confirm this is to use
an RF451 radio (
) connected to your lab computer.
The RF451 operates as the master radio device and the base station. The
CRVW3-RF451 device is configured as a slave device. (See Section
(p. 68)
, for more details about master/slave
radio configurations and base station considerations.)
Summary of Contents for CRVW3