Section 5. Overview
76
Calibration increases accuracy of a measurement device by adjusting its output, or
the measurement of its output, to match independently verified quantities.
Adjusting sensor output directly is preferred, but not always possible or practical.
By adding
FieldCal()
or
FieldCalStrain()
instructions to the CR800 CRBasic
program, measurements of a linear sensor can be adjusted by modifying the
programmed multiplier and offset applied to the measurement without modifying
or recompiling the CRBasic program.
5.2.8 Cabling Effects — Overview
Related Topics:
•
Cabling Effects — Overview
(p. 76)
•
Cabling Effects — Details
(p. 386)
Sensor cabling can have significant effects on sensor response and accuracy. This
is usually only a concern with sensors acquired from manufacturers other than
Campbell Scientific. Campbell Scientific sensors are engineered for optimal
performance with factory-installed cables.
5.2.9 Synchronizing Measurements — Overview
Related Topics:
•
Synchronizing Measurements — Overview
(p. 76)
•
Synchronizing Measurements — Details
(p. 387)
5.2.9.1 Synchronizing Measurements in the CR800 — Overview
5.2.9.2 Synchronizing Measurements in a Datalogger Network — Overview
Large numbers of sensors, cable length restrictions, or long distances between
measurement sites may require use of multiple CR800s.
5.3
Data Retrieval and Comms — Overview
Related Topics:
•
Data Retrieval and Comms — Quickstart
(p. 38)
•
Data Retrieval and Comms — Overview
(p. 76)
•
Data Retrieval and Comms — Details
(p. 426)
•
Data Retrieval and Comms Peripherals — Lists
(p. 568)
The CR800 communicates with external devices to receive programs, send data,
or join a network. Data are usually moved through a comms link consisting of
hardware and a protocol using Campbell Scientific
datalogger support software
(p.
Data can also be shuttled with external memory such as a
or a Campbell
Scientific mass storage media (USB: drive) to the PC.
Summary of Contents for CR800 Series
Page 2: ......
Page 4: ......
Page 6: ......
Page 32: ......
Page 34: ......
Page 54: ......
Page 92: ......
Page 310: ......
Page 446: ...Section 8 Operation 446 8 11 2 Data Display FIGURE 100 CR1000KD Displaying Data ...
Page 448: ...Section 8 Operation 448 FIGURE 102 CR1000KD Real Time Custom ...
Page 449: ...Section 8 Operation 449 8 11 2 3 Final Storage Data FIGURE 103 CR1000KD Final Storage Data ...
Page 450: ...Section 8 Operation 450 8 11 3 Run Stop Program FIGURE 104 CR1000KD Run Stop Program ...
Page 452: ...Section 8 Operation 452 FIGURE 106 CR1000KD File Edit ...
Page 456: ......
Page 462: ......
Page 523: ...Section 11 Glossary 523 FIGURE 116 Relationships of Accuracy Precision and Resolution ...
Page 524: ......
Page 526: ......
Page 556: ......
Page 558: ......
Page 560: ......
Page 597: ......