Section 7. Installation
250
Analog Voltage Measurement: Cluster Burst
'This program makes 500 measurements of two single-ended channels at 500 Hz.
'Sample pattern is 1,2,1,2
. Measurement cycle is repeated every 1 Sec. The following
'programming features are key to making this application work:
'--PipelingMode enabled.
'--Measurement speed set as follows:
' Scan() Interval=1, Units=Sec.
' SubScan() SubInterval=
2
, Units=mSec, and Count=
500
.
'--Scan() BufferOption increased to 5.
PipeLineMode
Public
ClusterBurstSE(2)
DataTable
(ClusterBurstSEData,1,-1)
Sample
(2,ClusterBurstSE(),FP2)
EndTable
BeginProg
'Scan(Interval,Units,BufferOption,Count)
Scan
(1,Sec,5,0)
'SubScan(SubInterval,Units,Count)
SubScan
(2,mSec,500)
'VoltSE(Dest,Reps,Range,SEChan,MeasOff,SettlingTime,Integ,Mult,Offset)
VoltSe
(ClusterBurstSE(),2,mV20,1,False,100,100,1.0,0)
CallTable
ClusterBurstSEData
NextSubScan
NextScan
EndProg
Summary of Contents for CR3000 Micrologger
Page 2: ......
Page 3: ......
Page 4: ......
Page 6: ......
Page 30: ......
Page 34: ......
Page 36: ......
Page 96: ......
Page 485: ...Section 8 Operation 485 8 11 2 Data Display FIGURE 110 Keyboard and Display Displaying Data ...
Page 487: ...Section 8 Operation 487 FIGURE 112 CR1000KD Real Time Custom ...
Page 491: ...Section 8 Operation 491 FIGURE 116 Keyboard and Display File Edit ...
Page 496: ......
Page 502: ......
Page 564: ...Section 11 Glossary 564 FIGURE 126 Relationships of Accuracy Precision and Resolution ...
Page 566: ......
Page 594: ......
Page 598: ......
Page 600: ......
Page 602: ......
Page 624: ......
Page 642: ......
Page 643: ......