The four radiation outputs can be measured using differential or single-ended
inputs on the datalogger. A differential voltage measurement is recommended
because it has better noise rejection than a single-ended measurement.
When differential inputs are used, jumper the low side of the input
to AG or
to keep the signal in common mode range.
The Tables 6-1 and 6-2 show the wiring instructions for the differential
measurement and single-ended measurement connections to the datalogger,
respectively. The cables have the white band at the pigtail end of the cable with
the colour keys. See the Figure 6-3 and 6-4 below for the labels on the cable for
both the SOLAR and TEMP cables.
Figure 6-3. Labels on the pigtail end of the SOLAR cable.
Figure 6-4. Labels on the pigtail end of the TEMP cable.