![Campbell CNR2 Instruction Manual Download Page 13](http://html1.mh-extra.com/html/campbell/cnr2/cnr2_instruction-manual_489147013.webp)
CNR2 Net Radiometer
Minimum(1,Net_Longwave,
IEEE4,False,True)
Maximum(1,Net_Longwave,
IEEE4,False,True)
Average(1,Net_Longwave,
IEEE4,False)
Average(1,Net_Rad,
IEEE4,False)
EndTable
'Main Program
BeginProg
Scan(1,Sec,1,0)
'CNR2 Net Radiation Measurements
VoltDiff(Net_Shortwave,1,mV25,1,True,200,250,
CNR2_SW_CAL,0.0)
VoltDiff(Net_Longwave,1,mV25,2,True,0,250,
CNR2_LW_CAL,0.0)
Net_Rad
=
Net_ShNet_Longwave
'Call Data Tables and Store Data
CallTable(Table1)
NextScan
EndProg
6.2.2 Example Program for CR10X Datalogger
;CNR2 Net Radiometer program for CR10X datalogger
;*** Wiring ***
; 1H Shortwave signal (white)
; 1L Shortwave signal reference (blue)
; 2H Longwave signal (brown)
; 2L Longwave signalreference- (black)
;gnd Shiled
(clear)
;
;*** Wiring ***
; 1H Shortwave signal (white)
; 1L Shortwave signal reference (blue)
;gnd
; 2H Longwave signal (brown)
; 2L Longwave signal reference (black)
;gnd Shield
(clear)
;
*Table 1 Program
01: 1.0000 Execution
Interval
(seconds)
; Measure Net Shortwave Radiation
1: Volt (Diff) (P2)
1: 1
Reps
2: 23
25 mV 60 Hz Rejection Range
3: 1
DIFF
Channel
4: 1
Loc [ Net_SW ]
5: 45.46
Multiplier
6: 0.0
Offset
9
Summary of Contents for CNR2
Page 18: ...CNR2 Net Radiometer 14 ...
Page 19: ......