CNR1 Net Radiometer
Wiring for Program Example 4
Color
Function
Example CR23X Program
Channels Used
Red CM3
Up
Signal
SE1
Blue
CM3 Up Reference
White
CM3 Down Signal
SE2
Black
CM3 Down Reference
Grey or Orange
CG3 Up Signal
SE3
Yellow CG3
Up
Reference
Brown
CG3 Down Signal
SE4
Green CG3
Down
Reference
Shield Shield
Pt-100 Temperature Sensor Connections to 4WPB100 and Datalogger
Color
Function
4WPB100
CR23X
H
5H
Red
Pt-100 Exci
L
5L
Blue
Pt-100 Excitation -
G
Yellow Pt-100
Signal
+
6H
Green Pt-100
Signal
-
6L
;{CR23X}
;
;CNR1 sensitivity for program example = 7.30 uV/W/m^2
;Multiplier for measurement instructions = 1000/7.3 = 136.99
*Table 1 Program
01: 2
Execution
Interval
(seconds)
;Measure CM3 Up and CM3 Down (shortwave radiation)
;Note: Multiplier (Parameter 6) will be different for each CNR1
1: Volt (SE) (P1)
1: 2
Reps
2: 22
50 mV, 60 Hz Reject, Slow Range
3: 1
SE
Channel
4: 1
Loc [ CM3_up ]
5: 136.99
Mult ; mult = 1000/CNR1 sensitivity
6: 0
Offset
;Measure CG3 Up and CG3 Down (longwave radiation)
;Note: Multiplier (Parameter 6) will be different for each CNR1
2: Volt (SE) (P1)
1: 2
Reps
2: 21
10 mV, 60 Hz Reject, Slow Range
3: 3
SE
Channel
4: 3
Loc [ CG3_up ]
5: 136.99
Mult ; mult = 1000/CNR1 sensitivity
6: 0
Offset
21
Summary of Contents for CNR1
Page 38: ...Appendix C CR3000 CR5000 Program that Controls the Heater C 6 ...
Page 39: ......