CMP6-L, CMP10-L, CMP11-L, and CMP21-L Pyranometers
22
;{CR10X}
*Table 1 Program
01: 10.0000 Execution
Interval
(seconds)
; CMP6 measurement in W/m2
1: Volt (Diff) (P2)
1: 1
Reps
2: 23
25 mV 60 Hz Rejection Range
;use the 50 mV range for the CR7, 21X and CR23X
3: 1
DIFF
Channel
;use the 250 mV range for the CR10X if
4: 3
Loc
[
Solar_Wm2
]
;calibration factor is > 25
μ
V/Wm-2
5: 69.7837 Multiplier
;1000/14.33
6: 0
Offset
; Set negative values to zero
2: If (X<=>F) (P89)
1: 3
X
Loc
[
Solar_Wm2
]
2: 4
<
3: 0
F
4: 30
Then
Do
3: Z=F x 10^n (P30)
1: 0
F
2: 0
n,
Exponent
of
10
3: 3
Z
Loc
[
Solar_Wm2
]
4: End (P95)
; Calculate units in MJ, where MJ = m * t * 0.000001.
; m = Solar_Wm2 from above, and t = 10 (scan interval).
5: Z=X*F (P37)
1: 3
X
Loc
[
Solar_Wm2
]
2: .00001 F
3: 4
Z
Loc
[
Solar_MJ
]
6: If time is (P92)
1: 0
Minutes
(Seconds
--)
into
a
2: 60
Interval
(same
units
as
above)
3: 10
Set Output Flag High (Flag 0)
7: Set Active Storage Area (P80)
1: 1
Final
Storage
Area
1
2: 101
Array
ID
8: Real Time (P77)
1: 1220
Year,Day,Hour/Minute
(midnight
=
2400)
9: Average (P71)
1: 1
Reps
2: 3
Loc
[
Solar_Wm2
]
10: If time is (P92)
1: 0
Minutes
(Seconds
--)
into
a
2: 1440
Interval
(same
units
as
above)
3: 10
Set
Output
Flag
High
(Flag
0)