Appendix B. Edlog Programming
Examples
B.1 Example CR10X Program
Instruction 5 (AC Half Bridge)
is used to measure the thermistor probe inside
the sphere.
Instruction 55 (Polynomial)
is used to calculate the temperature
in degrees Celsius. The polynomial coefficients are shown in TABLE B-1.
Thermistor resistance and computed temperature over a –10 to +84°C range is
shown in TABLE B-2. The example includes instructions for measuring an
HC2S3 to supply air temperature and relative humidity values. Calculations
for dewpoint, wet-bulb, and wet-bulb globe temperature are also included.
;{CR10X}
;Program: BLACKGLOBE.CSI
;Date: June 2013
*Table 1 Program
01: 5.0000 Execution
Interval
(seconds)
1: Batt Voltage (P10)
1: 1
Loc [ BattVolt ]
2: Internal Temperature (P17)
1: 2
Loc
[
CR10XTmpC
]
;Rotronic HC2S3 temperature and relative humidity.
;Sensor powered on all the time.
3: Volt (SE) (P1)
1: 1
Reps
2: 5
2500 mV Slow Range
3: 1
SE
Channel
4: 4
Loc
[
AirTempC
]
5: 0.1
Multiplier
6: -40
Offset
4: Volt (SE) (P1)
1: 1
Reps
2: 5
2500 mV Slow Range
3: 2
SE
Channel
4: 5
Loc [ AirRH ]
5: 0.1
Multiplier
6: 0
Offset
5: If (X<=>F) (P89)
1: 5
X Loc [ AirRH ]
2: 3
>=
3: 100
F
4: 30
Then
Do
B-1
Summary of Contents for BlackGlobe
Page 2: ......
Page 20: ...Appendix A The Theory of BlackGlobe Temperature and Heat Stress A 2...
Page 31: ......