Campbell ARG100 User Manual Download Page 17

User Guide 

3.  Check that the balance arm is free to move. This can be done by slowly     

pouring a measured quantity of water (say 25 cm

3

) through the gauge and 

counting the tips. It is worthwhile carrying this out at regular weekly intervals 
(for example, every Monday at 0900) while leaving the gauge connected to the 
datalogger. Providing a significant volume of water is used, these weekly 
checks can easily be identified in the logged measurements. This simple 
procedure confirms that the gauge is functioning, detects any marked change 
in the calibration and (if carried out punctually) introduces an independent 
time check into the records. 

6.  Calibration 

The sensitivity of the ARG100 is set at manufacture to approximately 0.2 mm/tip 
(optionally 0.25 mm/tip) and each gauge is subsequently calibrated as described 
in Section 6.1. The calibration factor is given on a certificate and is also recorded 
on the inside of the raingauge. For precise measurements use this value in your 
program instead of the nominal value of 0.2. 

6.1  Static Adjustment and Calibration 

The following procedure is carried out during manufacture and may be repeated if 
the calibration appears to have shifted: 

1.  Install the gauge over a sink unit as illustrated in Figure 2, ensuring that it is 

correctly levelled. 

2.  Adjust the two calibration screws under the tipping buckets until the balance 

arm tips in response to the correct amount of water dripped from a burette or 
pipette (10.13 cm

3

 for 0.2 mm sensitivity, 12.67 cm

for 0.25 mm).  

It is not possible to set the screws very precisely using this method, but it should 
be done with as much care as possible. It is obviously very important to ensure 
that both buckets tip in response to the same amount of water.  Many manufac-
turers and users of tipping bucket gauges aim to adjust the bucket settings until 
exactly the correct calibration is achieved. However, a dynamic test (see below) is 
required to check this calibration precisely after each readjustment and the process 
becomes very time-consuming. In any case, it is virtually impossible to get the 
adjustments absolutely correct, and it is generally preferable to adjust the settings 
as closely as is reasonably practical, and then to derive a calibration factor for 
each gauge individually. 

6.2  Dynamic Calibration 

1.  Set up the gauge as illustrated in Figure 2, carefully levelled and connected to 

the datalogger. 

2.  Fill the water container with 810.4 cm

3

 of water for a 0.2 mm calibration 

(1013.4 cm

3

 for a 0.25 mm setting). This is usually achieved most precisely 

and consistently by weighing the water on a balance capable of measuring to 
0.1 g (0.1 cm

3

). An alternative is to use a good quality graduated measuring 

cylinder.  

3.  Allow the water to drip slowly into the gauge (allowing approximately 100 

minutes for the container to empty). This is a rate equivalent to a rainfall of 
10 mm/hour, as recommended in BS7843 Section 2.1 for calibration 
purposes. At the end of this period approximately 80 tips will have occurred. 
The exact number is obtained from the datalogger, together with a visual 
inspection of what fraction of a ‘tip’ is left in whichever bucket is still filling 
as the flow of water finishes. 

Summary of Contents for ARG100

Page 1: ...ARG100 Tipping Bucket Raingauge User Guide Issued 30 11 10 Copyright 1998 2010 Campbell Scientific Ltd CSL 78...

Page 2: ......

Page 3: ...e repair is under guarantee or not Please state the faults as clearly as possible and if the product is out of the guarantee period it should be accompanied by a purchase order Quotations for repairs...

Page 4: ......

Page 5: ...l cell phones and aerials may also not be applicable according to your locality Some brackets shields and enclosure options including wiring are not sold as standard items in the European market in so...

Page 6: ......

Page 7: ...1 CR1000 Example Program 5 4 1 2 CR200 X Series Example Program 5 4 1 3 CR10X Example Program 6 4 2 Control Port Example 6 4 2 1 CR1000 Example Program 7 4 2 2 CR10X Example Program 8 5 Maintenance 8...

Page 8: ......

Page 9: ...ch causing contact to be made for a few milliseconds A two core cable is used to connect the gauge to the datalogger where the switch closures are counted The ARG100 is adjusted at manufacture to tip...

Page 10: ...may be more convenient as a base in which case through bolts or screws are suitable Whichever is chosen we recommend the use of large washers to spread the load more evenly around the mounting holes...

Page 11: ...talogger program the sensor should be wired to the channels shown on the wiring diagram created by Short Cut The ARG100 is typically wired to a datalogger s pulse channel see Table 3 1 Table 3 1 Wirin...

Page 12: ...bell Scientific s Short Cut Program Builder software You do not need to read this section to use Short Cut Precipitation is measured using a Pulse Count with a switch closure configuration code The mu...

Page 13: ...2 0 CallTable Rain NextScan EndProg 4 1 2 CR200 X Series Example Program CR200 X Series ARG100 Tipping Blk P_SW Clr ground Declare Variables and Units Public Rain_mm Units Rain_mm mm Define Data Tabl...

Page 14: ...utput Flag High Flag 0 3 Set Active Storage Area P80 1 1 Final Storage Area 1 2 101 Array ID 4 Real Time P77 1 1220 Year Day Hour Minute midnight 2400 5 Totalize P72 1 1 Reps 2 3 Loc Rain_mm Table 2 P...

Page 15: ...Table Define Data Tables DataTable OneDay True 1 DataInterval 0 1440 Min 10 Minimum 1 BattV FP2 False False Totalize 1 Rain_mm FP2 False ARG100 tipping bucket EndTable Main Program BeginProg Scan 5 Se...

Page 16: ...erval This section should be executed every scan and not placed in a subroutine or conditional statement 5 Maintenance To ensure reliable and accurate measurements we recommend that the following chec...

Page 17: ...cm3 for 0 25 mm It is not possible to set the screws very precisely using this method but it should be done with as much care as possible It is obviously very important to ensure that both buckets ti...

Page 18: ...terval is obtained by multiplying the number of recorded tips by the calibration factor This type of dynamic calibration gives repeatable results indoors but it is not a true representation of the gau...

Page 19: ...ations A tipping bucket gauge responds to discrete quanta of rainfall and the accuracy and reproducibility of this quantum are determined not only by factors such as friction in the bearings etc but a...

Page 20: ......

Page 21: ...lush with the bottom surface Tighten the nut up to the baseplate to lock the studs in position 4 Screw on the lower plastic thumb nuts and slide the raingauge over the studs until it is sitting on top...

Page 22: ......

Page 23: ...________ Calibration Factor _____________________________ Date _____________________________ Signed __________________________________________________________________ Campbell Scientific Ltd Campbell...

Page 24: ......

Page 25: ......

Page 26: ...rta T5M 1W7 CANADA www campbellsci ca dataloggers campbellsci ca Campbell Scientific Centro Caribe S A CSCC 300N Cementerio Edificio Breller Santo Domingo Heredia 40305 COSTA RICA www campbellsci cc i...

Reviews: