AM16/32 Relay Analog Multiplexer
25
21x
CR7
CR10(X)
CR23X/
CR5000
"4 X 16" Mode
E
E
H
H
AG
G
E/VX
H
L
L
L
COM H (ODD)
COM L
COM H (EVEN)
COM L
ODD H
ODD L
EVEN H
EVEN L
SHIELD
SENSOR SHIELDS
COM
MUXSIGNAL
SHIELD
FIGURE 13. Potentiometer Hook-up and Measurement
6.3.2 Potentiometer Measurement
Sensor to Multiplexer wiring – if panel switch is set to ”4X16” mode,
up to two potentiometers may be connected to one input SET.
Excitation and ground leads may be common; signal leads must be
routed separately (Figure 13).
Multiplexer to Datalogger wiring - Signal lines from two COM
terminals are connected to two consecutive single-ended analog input
channels. One COM terminal is connected to a datalogger switched
excitation channel, and the remaining COM line connects to
datalogger ground. Up to 32 potentiometers may be measured by two
single-ended datalogger channels.
6.3.3 Four Wire Half Bridge (Measured Excitation Current)
Sensor to Multiplexer Wiring - one sensor per input SET. The panel
switch is set to “4X16” mode.
Multiplexer to Datalogger Wiring - One COM line is tied to a
datalogger excitation channel, and two COM lines to a differential
analog input. The remaining COM line is connected to the H side of a
datalogger differential channel along with a fixed resistor. The other
side of the resistor connects to the L side of the differential channel
and to ground (Figure 14). Up to 16 four wire half-bridges may be
measured by two differential datalogger channels in this manner.