Campbell 237 Product Manual Download Page 9

237 Leaf Wetness Sensor 

and you will be prompted to send the program just created to the 

datalogger. 

7. If the sensor is connected to the datalogger, as shown in the wiring

diagram in step 4, check the output of the sensor in the datalogger support

software data display to make sure it is making reasonable measurements.

5. Overview

The 237 is often used for plant pathology applications because plant diseases 

are often associated with wet leaves. Duration of wetness and air temperature 

during wetness are inputs to many disease models.  

The sensor emulates a leaf, thereby approximating the wetness state of 

surrounding foliage. To determine leaf wetness, the 237 uses a circuit board 

with interlacing gold-plated fingers. Condensation on the sensor lowers the 

resistance between the fingers, which is measured by the datalogger. Droplets 

must touch two fingers simultaneously to change the sensor resistance. For this 

reason, the 237 is typically coated with flat latex paint, which spreads water 

droplets. 

The colo

u

r and type of paint affect sensor performance. Campbell Scientific 

supplies the sensor unpainted because individual modifications will vary with 

the application. The paper listed in Section 

10

References

 

(p. 9)

, describes the 

effects of paint colo

u

r and sensor angle on sensors of this type. 

The resistance of the sensor at the wet/dry transition point should be 

determined. A sharp change in resistance occurs in the wet-dry transition on 

the uncoated sensor; normally the transition is between 50 and 200 

kΩ

. Coated 

sensors have a poorly defined transition which normally occurs from 20 

kΩ

 to 

above 1,000 

kΩ

. For best results, the leaf wetness sensor should be field 

calibrated since the transition point will vary for different areas and vegetation. 

Features: 

Sensor is shipped unpainted so customer can choose appropriate

surface finish that best fits their application

Imitates characteristics of a leaf

Compatible with Campbell Scientific CRBasic dataloggers:

CR200(X) series, CR300 series, CR6 series, CR800 series, CR1000X,

CR1000, CR3000, CR5000, and CR9000(X)

6. Specifications

Resistance at Wet/Dry 

Transition: 

normally 50 and 200 k

 (uncoated sensor); 

normally 20 to 1000 k

 (coated sensor) 

Operational Temperature 

Range: 

0 to 100 °C 

Short-Term Survivability 

Temperature Range: 

–40 to 150 °C; sensor may crack when

temperature drops below –40 °C

Summary of Contents for 237

Page 1: ...237 Leaf Wetness Sensor Revision 01 2021 Copyright 1988 2021 Campbell Scientific CSL I D 506...

Page 2: ...under guarantee or not Please state the faults as clearly as possible and if the product is out of the guarantee period it should be accompanied by a purchase order Quotations for repairs can be give...

Page 3: ...cell phones and aerials may also not be applicable according to your locality Some brackets shields and enclosure options including wiring are not sold as standard items in the European market in some...

Page 4: ...beginning work Wear a hardhat and eye protection and take other appropriate safety precautions while working on or around tripods and towers Do not climb tripods or towers at any time and prohibit cl...

Page 5: ...nsor Preparation 4 7 2 Sensor Deployment 4 7 3 Wiring to Datalogger 5 7 4 Datalogger Programming 6 7 4 1 Measurement of Vs Vx 6 7 4 2 Calculating Sensor Resistance 7 8 Operation 7 8 1 Calculating Wet...

Page 6: ...Table of Contents ii 7 2 Measurement Instructions Parameters Results 7 9 1 Symptom Causes 9 CRBasic Examples B 1 CR1000 Program Example B 1 B 2 CR200 X Program Example B 1...

Page 7: ...rubber This compound was chosen for its resistance to temperature extremes moisture and UV degradation However this jacket will support combustion in air It is rated as slow burning when tested accor...

Page 8: ...ocessing will be required to determine parameters such as the percent of time wet Section 8 1 Calculating Wet Time Fraction p 8 4 After selecting the sensor click Wiring Diagram to see how the sensor...

Page 9: ...Scientific supplies the sensor unpainted because individual modifications will vary with the application The paper listed in Section 10 References p 9 describes the effects of paint colour and sensor...

Page 10: ...in plant disease and other applications Besides providing some protection for the gold plated electrodes this paint allows tiny water droplets to spread and bridge the electrodes Gillespie and Kidd1...

Page 11: ...uit schematic of the 237 TABLE 7 1 describes wiring to Campbell Scientific dataloggers FIGURE 7 2 237 Sensor Schematic TABLE 7 1 Wire Colour Function and Datalogger Connection Wire Colour Wire Functio...

Page 12: ...ndix A Importing Short Cut Code Into CRBasic Editor p A 1 Programming basics for CRBasic dataloggers are in the following section Complete program examples for select CRBasic dataloggers can be found...

Page 13: ...pressed in k is calculated as follows Rs R2 Vs Vx R2 R1 Therefore Rs k 1 Vs Vx 101 Except for the CR200 X Campbell Scientific also suggests including the following after the resistance calculation If...

Page 14: ...er to Appendix B Example Programs p B 1 for information on calculating fraction of time wet with the CR200 X 8 2 Calibration A wet dry threshold of 150 k is used in Appendix B Example Programs p B 1 W...

Page 15: ...ly cut if the mV200C mV25C or mV50C range code was used for the BrHalf instruction Another cause of these outputs is a highly conductive state 25 mV that occurs if the sensor is very wet with very ion...

Page 16: ...can now be edited for additional refinement Once the file is edited with CRBasic Editor Short Cut can no longer be used to edit the program it created 2 To add the Short Cut wiring information into t...

Page 17: ...V25 1 VX1 1 2500 True 0 250 1 0 Rs_kOhms 1 Vs_Vx 101 If Rs_kOhms 10000 Then Rs_kOhms 10000 EndIf CallTable Wetness NextScan EndProg CRBasic Example B 2 CR200 X Program Example CR200 X Series Datalogge...

Page 18: ...lculate Time Fraction Wet at top of each hour If IfTime 0 60 Min Then Interval must match data table interval TimeFractionWet ScanIntervalWet ScanIntervalSum ScanIntervalWet 0 ScanIntervalSum 0 EndIf...

Page 19: ...nfo campbellsci fr www campbellsci fr Germany Location Phone Email Website Bremen Germany 49 0 421 460974 0 info campbellsci de www campbellsci de India Location Phone Email Website New Delhi DL India...

Reviews: