![Campbell 237 Product Manual Download Page 15](http://html1.mh-extra.com/html/campbell/237/237_product-manual_3761195015.webp)
237 Leaf Wetness Sensor
9
9. Troubleshooting
lists the causes of unusual resistance readings.
TABLE 9-1. Symptom/Causes
Resistive Value
Causes
IEEE4
FP2
INF,
≥
99999,
≤
-99999
INF, ±7999
The 1 k
Ω
bridge resistor holds the input channel at 0 mV when the
sensor is completely dry. However, the measurement may intermittently
deviate from zero slightly, but still be within the resolution
specifications of the datalogger. When this occurs, Rs = either a very
large or a very small number.
NAN, -100,
-INF
NAN
NAN can indicate that no sensor is connected or a cable has been
cleanly cut if the
mV200C
,
mV25C
, or
mV50C
range code was used for
the
BrHalf
instruction.
Another cause of these outputs is a highly conductive state (> 25 mV)
that occurs if the sensor is very wet with very ionic water.
< 0
If the measured voltage exceeds 24.75 mV, but does not exceed the
input voltage range, the result of the bridge equation becomes negative.
10. References
1
Gillespie, T.J. and Kidd, G.E. 1978. Sensing duration of leaf moisture
retention using electrical impedance grids. Can. J. Plant Sci. 58:179-187.
2
East, David (Ohio State University). 1994. Field Testing of Phone Accessible
Multi-Channel Datalogger for Tomato IPM Programs. Unpublished.
The citation of researcher does not imply the endorsement of
Campbell Scientific products by any researcher or institution.
NOTE