229 Heat Dissipation Matric Water Potential Sensor
8: Excitation with Delay (P22)
;Wait 29 more seconds for next reading
1: 1
Ex
Channel
2: 0
Delay W/Ex (0.01 sec units)
3: 2900
Delay After Ex (0.01 sec units)
4: 0000
mV
Excitation
9: Thermocouple Temp (DIFF) (P14)
;Take 30 second temperature reading
1: 1
Reps
2: 1
2.5
mV
Slow
Range
3: 4
DIFF
Channel
4: 1
Type
T
(Copper-Constantan)
5: 1
Ref Temp (Deg. C) Loc [ ref_temp ]
6: 4
Loc [ T30sec_1 ]
7: 1.0
Mult
8: 0.0
Offset
10: Do (P86)
;turn off CE8
1: 52
Set
Port
2
Low
11: Z=X-Y (P35)
;Calculate delta T
1: 4
X Loc [ T30sec_1 ]
2: 3
Y Loc [ T1sec_1 ]
3: 5
Z Loc [ deltaT_1 ]
12: Do (P86)
;Save readings to final storage
1: 10
Set Output Flag High (Flag 0)
13: Set Active Storage Area (P80)
1: 1
Final
Storage
Area
1
2: 229
Array
ID
14: Real Time (P77)
1: 1220
Year,Day,Hour/Minute
(midnight
=
2400)
15: Sample (P70)
1: 1
Reps
2: 2
Loc [ Tinit_1 ]
16: Sample (P70)
1: 1
Reps
2: 5
Loc
[
deltaT_1
]
17: Do (P86)
;Disable flag 1
1: 21
Set
Flag
1
Low
18: End (P95)
15