background image

Model 109B & 109 BAM Temperature Probe 

The 60 and 50 Hz integration options include a 3 ms settling time; longer 
settling times can be entered into the Settling Time parameter.  The example 
Therm109 instruction listed below has a 20 mSec (20000 µSec) delay: 

‘Therm109 ( Dest, Reps, SEChan, ExChan, SettlingTime, Integ, Mult, Offset ) 

Therm109(T109B_C,1,1,1,20000,_60Hz,1.0,0.0) 

 

In Edlog use the DC Half Bridge instruction (P4) with a 20 millisecond delay 
as shown below.  Use P4 in place of P5 in Example 5.2.1.1 (the instructions 
that follow P5 to convert the measurement result to temperature are still 
required). 

1:  Excite-Delay (SE) (P4) 
 1:  1 

 Reps 

  2:  25 

  2500 mV 60 Hz Rejection Range (Delay must be zero) 

 3:  1 

 SE 

Channel 

  4:  1 

  Excite all reps w/Exchan 1 

  5:  2 

  Delay (0.01 sec units) 

 6:  2500 

 mV 

Excitation 

  7:  3 

  Loc [V_Vx   ] 

 8:  .0004 

 Multiplier 

 9:  0.0 

 Offset 

 

6.  Measurement Details 

Understanding the details in this section are not necessary for general 
operation of the 109B Probe with CSI's dataloggers. 

The Therm109 Instruction outputs a 2500 mV excitation and measures the 
voltage across the 24.9 K resistor (Figure 6-1).  The thermistor resistance 
changes with temperature.   

EX 

HI 

AG 

BLACK 

RED 

PURPLE 

CLEAR 

24.9K 

Ω

, 0.1% 

10K3AI 

THERMISTOR

 

FIGURE 6-1.  109B Thermistor Probe Schematic 

11 

Summary of Contents for 109B

Page 1: ...INSTRUCTION MANUAL 109B 109BAM Temperature Probe April 2010 Copyright 2010 Campbell Scientific Canada Corp...

Page 2: ...glect accidents of nature or shipping damage This warranty is in lieu of all other warranties expressed or implied including warranties of merchantability or fitness for a particular purpose CSC is no...

Page 3: ...CRBasic Examples 6 5 1 1 1 Sample Program for CR200 X Series Datalogger 7 5 1 1 2 Sample Program for CR1000 Datalogger 7 5 1 1 3 Sample Program for CR1000 Datalogger AM16 32B 4x16 mode 8 5 2 Edlog 8 5...

Page 4: ...each single ended channel used This will vary depending on the multiplexer configuration and the number of sensors measured The model name 109B will be used in this manual to refer to both 109B and 10...

Page 5: ...tion in air It is rated as slow burning when tested according to U L 94 H B and will pass FMVSS302 Local fire codes may preclude its use inside buildings NOTE 2 Accuracy The overall probe accuracy is...

Page 6: ...rature Degrees C Possible Error Degrees C Thermistor Tolerance CR200 Bridge Measurement Error 0 06 of reading 2 4 mV CR10X Bridge Measurement Error CR200 Resolution 0 5 mV 0 6 mV 24 9 k Ohm Fixed Resi...

Page 7: ...physically limited by the number of lead wires that can be inserted into a single voltage excitation terminal approximately six TABLE 4 1 109B Connections to Campbell Scientific Dataloggers Colour Hea...

Page 8: ...COM ODD H Vx 1 E 1 Vx 1 Signal Return 1 White COM ODD L SE 1 SE 1 SE 1 Signal Return 2 Black COM EVEN H SE 2 SE 2 SE 2 Signal Return 3 Green COM EVEN L SE 3 SE 3 SE 3 Shield Clear COM G The wiring ass...

Page 9: ...gths prior to programming your datalogger Measurement details are provided in Section 6 5 1 CRBasic In the CR200 X series CR800 CR850 CR1000 and CR3000 dataloggers Instruction Therm109 is used to meas...

Page 10: ...terval 0 10 min Average 1 Air_Temp 0 EndTable BeginProg Scan 1 sec Measure the temperature Therm109 Air_Temp 1 1 Ex1 1 0 0 Call the data table CallTable AvgTemp NextScan EndProg 5 1 1 2 Sample Program...

Page 11: ...9BAM_C 1 3 1 1 0 _60Hz 1 0 Switch to next AM16 32 Multiplexer channel set PulsePort 1 10000 109BAM Temperature Probe 3 wire measurements T109BAM_C 4 5 6 on the AM16 32B Therm109 T109BAM_C 4 3 1 1 0 _6...

Page 12: ...ution Interval seconds 1 AC Half Bridge P5 1 1 Reps 2 25 2500 mV 60 Hz Rejection Range 3 1 SE Channel 4 1 Excite all reps w Exchan 1 5 2500 mV Excitation 6 1 Loc V_Vx 7 1 0 Mult 8 0 0 Offset 2 Z 1 X P...

Page 13: ...2 9 Loc Air_Temp Table 2 Program 02 0 0000 Execution Interval seconds Table 3 Subroutines End Program 5 3 Electrically Noisy Environments AC power lines pumps and motors can be the source of electric...

Page 14: ...w P5 to convert the measurement result to temperature are still required 1 Excite Delay SE P4 1 1 Reps 2 25 2500 mV 60 Hz Rejection Range Delay must be zero 3 1 SE Channel 4 1 Excite all reps w Exchan...

Page 15: ...trusion For all factory repairs and recalibrations customers must get a returned material authorization RMA Customers must also properly fill out a Declaration of Hazardous Material and Decontaminatio...

Page 16: ...5 4 Make sure the clear shield wire is connected to datalogger ground and the datalogger is properly grounded For all factory repairs customers must get an RMA Customers must also properly fill out a...

Reviews: