Appendix A. Analog Output
Configurations
The 092 ships from Campbell Scientific pre-wired and pre-configured for
analog output with the following settings:
•
Analog Output Voltage: 0 – 5 Vdc
•
Range: 600 – 1100 mbar
Output voltage and measurement ranges are set by DIP switches. Dip switch
settings are listed in Tables A-1 and A-2. Highlighted settings are those set at
the factory by Campbell Scientific, Inc. Complete information concerning
measurement, output, and communications settings is available in the Met One
092, 594, 6633A Barometric Pressure Sensor Manual
, which is available at
TABLE A-1. Dip Switches SW1:
Analog Output Voltage Settings
Range SW1-1
SW1-2
0-1 Vdc
On
On
0-2 Vdc
On
Off
0-2.5 Vdc
Off
On
0-5 Vdc*
Off*
Off*
*Factory configuration at Campbell Scientific, Inc.