PM 1001937 000 00
Operating manual SIRAX BT7300 / BT7350
22/28
Now select whether you want to save the configuration in a file or
transfer it to the device.
This screen allows you to check the functionality of the device.
7.2. Programming over Modbus Adress-Register
Modbus
Adress
Register Name
Comment
Register Type
Read /
Write
Default
values
Range
40001
Maschin_ID
ID Machine
Unsigned short
R
40002
FW_Version
Firmware Release
Unsigned short
R
0
40003
Adress
Modbus Adress
Unsigned short
R/W
1
1 ... 250
40004
Delay
Answer Delay
Unsigned short
R/W
1
1 ... 1000
40005
Baudrate
Baudrate: 0=1200 / 1=2400 / 2=4800 /
3=9600 / 4=19200 / 5=38400 / 6=57600
/ 7=115200
Unsigned short
R/W
1
0 ... 7
40006
Parity
0=NO / 1=ODD / 2=EVEN
Unsigned short
R/W
0
0 ... 2
40007
DC Filter
Number of tenths of second (1/10) for all RMS
calculation in DC
Unsigned short
R/W
10
1 ... 65535
40008
Flag
Measurement
Bit 0:
Bit 1:
Bit 2:
0=TRMS value (without sign);
1=DC measurement (with sign)
0=Energy storing disable;
1=Energy storing enable
0=Frequency detect on voltage channel;
1=Frequency detect on current channe
Unsigned short
R/W
0x10
40009
TV Ratio
Voltage transformer ratio
Float (LSW first)
R/W
1.0
40010
40011
TA Verhältnis
Stromwandlerverhältnis
Float (LSW first)
R/W
1.0
40012
40013
Strom-/Leistungs-
abschaltung
LSB:
Strom in mA (250)
MSB:
Leistung in W (1)
Unsigned short
R/W
siehe Be-
schreibung
40014
# von ZX für die VI
Messung
Anzahl der ZX für die AC-Messnummer des Null-
durchgangs des Linienzyklus für die AC-Messung
RMS
Unsigned short
R/W
50
1 ... 65535