C3 Specifications
Measuring ranges
30A / 300A / 3000A AC
Output sensitivity
100mV / 10mV / 1mV per A (AC coupled)
Accuracy (45-65Hz)
± 1% of reading ± 0.1A 30/300A
± 1% of reading ± 1A 3000A
Minimum load
100k for specified accuracy
Frequency range
10Hz to 20kHz (-1dB)
Phase
<± 1° (45-65 Hz)
Position sensitivity
± 2% of reading
External field
± 0.2% of range with cable >200mm (8") from the probe
Noise
8mV eff. in 30A range
2mV eff. in 300A / 3000A ranges
Temperature coefficient
± 0.1% of reading /°C
Power supply
Provided by CAM mobile
Overload
Indicated by a red LED
Enclosure
Material
ARNITE T06-200SNF, UL94 V0
Degree of Protection
IP40
Dimensions
110 (l) x 65 (w) x 23 (d) mm
Probe
Probe length
610mm (24”), double insulated
Probe diameter
9.9 mm (0,39”)
Output cable
2m (78.7“),probe to integrator
Material
Alcryn 2070NC, LATI LATENE 7H2W V0
Degree of Protection
IP65
General Characteristics
Operating temp.
-20°C up to +80°C
Storage temp.
-40°C up to 90°C
Operating humidity
15% up to 85% (non-condensing)
Safety standards
EN 61010-1:2001
EN 61010-031:2002
EN 61010-2-032:2002
EMC
EN 61326-2-2 :2006
1000 V
RMS
, Category III, 600 V
RMS
, Category IV, Pollution Degree 2 (Probe and integrator)
30V maximum between output and earth (ACP 3005)
ROHS and WEEE compliant
Rated for continuous use