A-5
MC-III™ Panel Mount Flow Analyzer
Appendix A
setting is recommended when the software is connecting directly to the instrument. The fixed timeout type is
recommended when data is being transmitted through a radio or other network device (serial to TCP/IP) and
there are recognizable delays in the communication system.
Timeout Period
When a fixed timeout period type is selected, the user enters the timeout period value (in milliseconds) in the
Timeout Period field. When an “optimal” timeout period type is selected, the user can use this field to enter
a percentage by which the “optimal” timeout period is extended. This extension can be advantageous when
connecting to a slow computer.
Downloading Options
The MC-III Panel Mount automatically saves daily flow logs on the contract hour, and hourly flow logs
around the clock. The instrument also automatically saves an event log each time a user change is made. By
default, the logs are saved as .csv files.
The Downloading Options screen allows a user to change the format of the auto-save flow logs and event logs
to Excel (.xls) files. To save the logs in .xls format, check the “XLS-MS Excel File” checkbox (Figure A.5).
Additional checkboxes allow for changes in the appearance of downloaded log data.
Users can also export flow data to a .csv or .xls file as needed. For complete information, see
Figure A.5—Menu for changing the auto-save log file format
Summary of Contents for NUFLO MC-III
Page 1: ...NUFLO TM MC III Panel Mount Flow Analyzer User Manual Manual No 2350326 01 Rev 02 ...
Page 36: ...36 Section 3 MC III Panel Mount Flow Analyzer ...
Page 66: ...66 Section 4 MC III Panel Mount Flow Analyzer Figure 4 30 4 20 mA output test screen ...
Page 86: ...86 Section 5 MC III Panel Mount Flow Analyzer ...
Page 100: ...B 2 Appendix B MC III Panel Mount Flow Analyzer ...
Page 118: ...C 18 Appendix C MC III Panel Mount Flow Analyzer ...
Page 119: ...C 19 MC III Panel Mount Flow Analyzer Appendix C ...