7

CD-20SFA Sample Probe Controller 

 

Table 1.1—CD-20SFA Specifications

24 VDC

115/230 VAC

Temperature

-40ºC to 55ºC

-40ºC to 55ºC

Voltage

24VDC @2A max

115 / 230 VAC @ 1A max

Pulse Input

12VDC

12VDC

Transistor Output

12VDC max @ 100mA max

12VDC max @ 100mA max

Current Draw (Controller)

Stop Mode

30 mA

60 mA

Running Mode

0.5 Amp

150 mA

Peak Motor (Turn On) Current

1 Amp

300 mA

Current Draw (Motor)

Motor (Turn On) Current

320 mA

850 mA

Section 2—Installation and Wiring

!

  WARNING: This module is designed for connection to hazardous electric voltages. Ignoring 

this warning can result in severe personal injury or mechanical damage. This product must 

be reliably earthed and installed by qualified personnel in accordance with the prevailing local 

electrical wiring regulations and safety standards.

The CD-20SFA controller mounts directly to the C-Series sample probe via the DC-2 coupling (Figure 2.1). 

The DC-2 coupling should be hand-tightened to ensure that the internal retaining ring is not pushed out of 

its groove. If the DC-2 coupling is over-tightened at the CD-20SFA hub, the motor shaft will not engage the 

DC-1 coupling on the probe.

Figure 2.1—DC-2 coupling 

All field connections are made to terminal block 1TB on the control card (

Figure 2.2, page 

8

). Install 

wiring as follows:

1. 

Connect power to 1TB-1 and 1TB-2.

2. 

Connect the flow pulse input contact to 1TB-5 and 1TB-6. If an open collector transistor driver is used, 

connect the collector to 1TB-5 and connect the emitter to 1TB-6 (common).

Summary of Contents for CLIF MOCK CD-20SFA

Page 1: ...CLIF MOCK CD 20SFA Sample Probe Controller User Manual Manual No 2350754 01 Rev 02 Class I Groups C and D Hazardous Locations...

Page 2: ...d within the parameters stated in this manual and in conjunction with the ratings printed on the product serial tag Technical Support Contact Information Cameron Measurement Systems Division 14450 Joh...

Page 3: ...5 How It Works 6 Hazardous Location Safety Compliance 6 Table 1 1 CD 20SFA Specifications 7 Section 2 Installation and Wiring 7 Section 3 Maintenance 8 Operational Test 8 Basic CD 20SFA Check Out 9 S...

Page 4: ...ler s liability and Buyer s exclusive remedy in any case of action whether in contract tort breach of warranty or otherwise arising out of the sale or use of any products software supplies or material...

Page 5: ...control card potentiometer Alarm delays can range from 5 seconds to 6 minutes An explosion proof housing encloses the complete assembly and two in NPT holes allow connection of the line power and inpu...

Page 6: ...rols the speed of the motor at about 12 rpm and current feedback limits motor current to 3 amperes The pulse input be closed for at least 5 msec to produce a start pulse An electronic square wave inpu...

Page 7: ...bly earthed and installed by qualified personnel in accordance with the prevailing local electrical wiring regulations and safety standards The CD 20SFA controller mounts directly to the C Series samp...

Page 8: ...failure alarm relay contact to 1TB 3 and 1TB 4 The contact will remain open for as long as sampling occurs within the alarm delay period This contact is rated for 90 VA resistive and 30 VA inductive...

Page 9: ...to 1TB 5 and 1TB 6 2 Mount the CD 20SFA to the sample probe and hand tighten the DC 2 coupling 3 Verify that the product line is full and pressurized 4 Reconnect input power to the CD 20SFA 5 Hold a...

Page 10: ...D1 a 33V crowbar zener diode pro vides over voltage protection for the rest of the circuit Diode D2 prevents current flow if VDC power is connected to the common terminal and a ground is connected to...

Page 11: ...mity switch is not sensing the high points of the cam Check proximity switch and cam alignment CD 20SFA control failure power supply board Return to factory for replacement of CD 20SFA motor control a...

Page 12: ...R U S S I A 603 5569 0501 ms kl c a m com...

Reviews: