LiveTrack AV for fMRI installation manual
Version R02
11
Configuring LiveTrack and interfaces
By default HID and UVC interface are turned on. It is not recommended to change this setting
but it is possible to set which USB interfaces are active by editing the
Firmware.xml
file located
on the LiveTrack device. You can edit this file with any suitable text editor (for instance Notepad
on Windows).
The rest of the settings are parameters for the tracking algorithm and should not be changed.
Once you have finished editing the LiveTrack.xml file, exit the text editor (saving the file when
prompted). The status LED will light up to indicate activity (the drive is rather slow, compared
to other USB drives you might use).
Once the status light has gone out
, eject the drive, then re-boot the LiveTrack box by
disconnecting the USB lead, waiting for 2 seconds, then reconnecting it. Do not press the push
button this time.
When editing, saving or deleting files on the LiveTrack system, an orange light next to the USB input will
be illuminated. While in Mass Storage Device-mode the device should only ever be unplugged or reset if
this light is off. Be sure to use the eject function in your operating system before disconnecting the device.
Any cycling of power while the light is illuminated could result in corrupted data.