P r e - O p e r a t i o n P r e p a r a t i o n
1. Verify
Luna HD
software and hardware setup.
2. Clean
Luna HD
(first time use) as necessary.
3. Install a new Camera Sheath on the handpiece for each patient.
IMPORTANT
Make sure to align the optically clear section of the sheath
with the camera lens by facing the camera/light source
section of the handpiece down toward the paper layer.
Camera Sheath Installation -
1 Insert the handpiece tip between the second
and third layers with the camera/light facing
towards the paper layer.
See A and B
2. Completely insert tip until it is fully into the
narrow section of the sheath.
3. Squeeze the illuminated tip to ensure that the
optically clear area of the Camera Sheath is
flat against the lens tip.
See C.
4. Peel away and discard the top plastic layer
and the bottom paper layer.
See D
.
Step 1 and 2
Step 4
Step 3
2
The Camera Sheath must be used for only one patient and disposed
of properly in accordance with local code.
Surface Cleaning Instructions:
Clean the handpiece following each patient use. The cleaning instructions must be followed carefully to prevent damage to the
camera housings, and internal components. Cleaning products containing active ingredients listed in the Product Care Guidelines are
prohibited for use with
Luna HD
. These chemicals can cause damage to the camera and could void the warranty.
6
5
1. Disconnect the USB cable. Remove the Camera Sheath and dispose in accordance with local code.
2. Moisten a cloth with warm water and remove any excessive moisture from the cloth before wiping the handpiece.
3. Wipe outer camera body with the moist cloth to ensure no pooling of water surrounds capture switch, ON/Standby button or USB
connection.
4. The window lens may be cleaned with a cotton swab which is moistened with 70% isopropyl alcohol.
5. Allow the handpiece sufficient time to air dry before applying a new Camera Sheath, or storing the camera prior to the next use.
Do not autoclave the Camera Handpiece
Do not apply excessive pressure when wiping; do not scrub.
Do not rinse or immerse the handpiece in liquids.
Do not spray the handpiece directly with liquids .
O p e r a t i n g P r o c e d u r e
Important: Patient Safety -
Make sure you are using a new Camera Sheath for each patient.
7
The PDF version of Operator’s Manual for Luna HD is included on the CD. Also included are USB Device Drivers, TWAIN
Interface, Demonstration Program, Software Installation and Configuration Guide and Adobe Acrobat Reader
Authorized imaging software, purchased from your dealer or other company, must be installed on the computer in order
to operate Luna HD.
Make sure that the computer meets all requirements necessary to support the Luna HD. See Minimum Computer System
Requirements in the Operator’s Manual.
1. Start the imaging software.
2. Press the ON/Standby button to activate camera.
3. Set the red dot of the focusing ring to the desired view setting mark.
The red dot may be adjusted between the marks as desired.
4. Position the camera lens window over the object to be imaged and
view the image on monitor. Make small adjustments to the focusing
ring position as needed to adjust image clarity.
5. Press and release the handpiece top switch to freeze the displayed
video on the monitor. If image is not suitable, proceed to step 6.
When image is suitable, perform to step 7.
6. Press and release the top switch again to unfreeze the displayed image
and return to a live video display. Return to step 5 to freeze video.
7. Press and release the handpiece bottom switch to save the frozen
image to the computer hard drive and return to a live video display.
8. Repeat steps 3 through 7 as necessary.
9. Press the ON/Standby button to turn the camera OFF.
ON/
Standby
Top
Button
Bottom
Button
a
d
b
c
Focusing
Ring
Handpiece
Connector
a. 50 to 300 mm Face - Smile - Arch
b. 15 mm
Intraoral Mouth Tour
c. 8 mm
Single Tooth Intraoral
d. 2 mm
Macro view
Focusing Ring Settings
b
c
d
a
Scan QR Code for
more about Luna HD.
Camera
Sheath
Camera
Handpiece
Light
Plastic
Layer 1
Plastic
Layer 2
Plastic
Layer 3
Squeeze
Plastic
Layer 1
Paper
Layer
A
B
C
D