C300 Transmission protocol for C300 firmware 5.x.x
C300B transmission protocol 2017-06-12.doc
13/22
4.3.1. Frequency measure with C300 calibrator as a reference source (example for number of measuring
impulses mode and 200 measuring impulses)
4.3.2. Frequency measure with external reference (example for measuring time mode and 15s measuring
time for both S0 inputs – first S0 input measure a frequency of electricity meter, second S0 input
measure a frequency of external reference)
START
Set parameters
of calibrator
outputs
Set first S0 input settings
to measure 200 impulses
WRMETS0_0,2,200
Set first S0 input mode
settings and start counting
WRMETS0_0,0,2
Frequency
≥0
YES
NO
Read measured
frequency
RDMETS0_0,4
Switch Off first S0
input
WRMETS0_0,0,0
START
Set parameters
of calibrator
outputs
Set first S0 input settings
to measure 15s
WRMETS0_0,2,15
Set second S0 input
settings to measure 15s
WRMETS0_1,2,15
Set first S0 input mode
settings and start counting
WRMETS0_0,0,1
Set second S0 input mode
settings and start counting
WRMETS0_1,0,1
Frequency
≥0
YES
NO
Read measured
frequency of first S0 input
RDMETS0_0,4
Switch Off first S0
input
WRMETS0_0,0,0
Frequency
≥0
YES
NO
Switch Off first S0
input
WRMETS0_0,0,0
Read measured frequency
of second S0 input
RDMETS0_1,4