Twister II - Maintenance 108
P/N 113564 Rev. 2
Twister II Robot User’s Manual
Caliper Life Sciences, Inc.
Changing the Assay Engine Twister II Fuses
Assay Engine Twister II Fuses:
•
100 - 240V~ operation:
2 x T2.5A250V, 5 x 20mm (P/N 114703)
NOTE
To order replacement fuses, please contact Caliper Technical
Support. See
.
The Assay Engine Twister II operates at 100 - 240V~ using two
2.5A fuses. For instructions for the Twister II, see
Twister II Fuse(s)” on page 105
.
WARNING
•
Electrical shock hazard. Disconnect the power cord before
changing the fuses.
•
For continued fire protection and correct functioning of the unit,
replace fuses only with exact part number to prevent fire.
To change the Assay Engine Twister II fuses:
1
Verify the Assay Engine Twister II power switch is OFF and the
power cord is unplugged.
2
Using a small flat-blade screwdriver at the top of the fuse
housing, gently pry the fuse housing (see
) out of the
power entry port.
Figure 87. Assay Engine Twister II Power Entry Port
Pry Here