INSTRUCTIONAL MANUAL
2750L-P/4500L-P/6000L-P
August, 2001
Rev. K
-120-
8.3.4 EMERGENCY
OPERATION
(Refer to MIL-704D Doc. 5.2.5)
1. Voltage per Figure 14, page 156
2. Frequency per Figure 15, page 156
8.4
KEYPAD ENTRY (Refer to page 31 for Keyboard flow chart)
To perform a test from the key board, the following key sequence is required:
704
ENT
The following screen will appear:
MIL704D:SelA,B,C
ENT=all
CLR=EXIT
Pressing the A, B, C or any combination selects the phase in test. Press ENT without the phase select for
simultaneous three-phase test.
The following screen appears for a short time.
TEST
A,B,C
CLR
to
Reselect
The next screen is:
Apply
Nom
Output
Press
ENT
When ENT is selected the following screen appears:
Press MNU to
Select
Test
The MNU screen has two lines of selection shown at a time.
There are 3 different types of operations that can be selected from a MENU acreen. If the word MENU appears for the
item selected, another MENU screen will be displayed. If the word TEST appears fir the item selected, the test will
start. The display will return to the previous screen if the word RETURN appears for the item selected.
The Main Menu will appear as follows:
1=Steady
St
Menu
2=Transient
Menu
3=Abnormal
Menu
4=Emergency
Menu
5=MIL704D
Test
6=Return
If key 1 is selected "Steady State: from the Main Menu, the following Menu will appear:
1=Voltage
Test
2=Unbalance
Test
3=Phase
dif
Test
4=Wave
dist