CALIBRATION
TW SERIES
3-28
Wait for the DMM to settle and record the value as
RDNG2
.
Calculate the gain and offset for the external input:
GAIN
= 60.0 / (
RDNG2
–
RDNG1
)
OFFSET
= 10 – (
RDNG2
/ (42.0 *
GAIN
))
Open the output relay and turn off the external gain port with the following GPIB
commands:
OUTPUT OFF
SYST:EXT:GAIN 0
Send the updated calibration constants to the TW with the following GPIB strings:
CAL:UNIQ:GEXTG
GAIN
CAL:UNIQ:OEXTG
OFFSET
3.19 LOW RANGE CURRENT CALIBRATION
Connect the 12 ohm resistor to the output of the TW through the power analyzer. Set
up the power analyzer to measure current.
Set up the TW to DC coupling, low range, local sense, 13 ampere current limit, output
relay closed and 12 volts with the following GPIB programming strings:
OUTP:COUP DC
SOUR:VOLT:RANGE LOW
SOUR:SENSE LOCAL
SOUR
{1,2,3}
:CURR 13.00
OUTPUT ON
SOUR
{1,2,3}
:VOLT:OFFSET 12
Wait until the power analyzer has stabilized and record the current reading as
RDNG1
.
Query the TW’s current using the GPIB string below, and record this value as
MEAS1
.
MEAS
{1,2,3}
:CURR?
Program the TW to 120 volts DC using the string below:
SOUR
{1,2,3}
:VOLT:OFFSET 120
Wait until the power analyzer has stabilized and record the current reading as
RDNG2
.
Query the TW’s current using the GPIB string below, and record this value as
MEAS2
.
MEAS
{1,2,3}
:CURR?
Summary of Contents for TrueWave TW1750
Page 2: ......
Page 4: ...ii This page intentionally left blank ...
Page 10: ...TABLE OF CONTENTS TW SERIES viii This page intentionally left blank ...
Page 36: ...MAINTENANCE AND TROUBLESHOOTING TW SERIES 2 16 This page intentionally left blank ...
Page 80: ...CALIBRATION TW SERIES 3 44 This page intentionally left blank ...