3. Menu description
Press and hold the
button for 3 seconds to access the menu. Buttons are used to navigate
through the items. Press the button to access desired position. You can choose from the
following menu items:
<Measurement time>
<View results>
<Backlight>
<Language>
<Calibration>
<System information>
<Default settings>
<Exit>
<Measurement time>
With this function you can set measurement time. Use the buttons to set desired hours. Press
the button to move to the next parameter (minutes), repeat this procedure until the desired
minutes and seconds have been set.
<View results>
This option allows you to see the last ten saved measurement results. Press the buttons to
scroll through the list of results. Press the button to go back to the menu.
<Backlight>
Press the buttons to increase/decrease the backlight level. Press the button to save the
setting and back to the menu.
<Language>
Use the buttons to switch between languages (English, Deutsch, Nederlands). Press to
save the setting and go back to the menu.
<Calibration>
Calibrating the Integrating Sound Level Meter (SL-200) requires an acoustical calibrator which
provides a stable sound pressure at 1kHz and 94dB.
Using the buttons, you can select one of the two types of calibration:
1. Factory – Select this option if you want to use the meter factory calibration set during
manufacture.
2. User – Select this option if you want to use your own calibration.
If you decide to select option "User," you will be able to choose between a new calibration (NEW –
press button ) and the calibration of the user stored in the device's memory (CURRENT – press
button ).
CAISSON Elektronik GmbH
Boschstrasse 16, D-47533 Kleve, Deutschland
tel: +49(0)2821-894390
F/S
F/S
A/C
To extend battery life during measurement, the meter automatically turns off the display backlight if you
do not press any button for 6 minutes. To turn the backlight on again, press any button.