12 - ENG    PN MN234 G | User Manual

Liberator

Table 2*

Recommended separation distances between portable and mobile 

RF communications equipment and the Liberator

The Liberator is intended for use in an electromagnetic environment in which radiated RF disturbances 

are controlled. The customer or the user of the Liberator can help prevent electromagnetic interfer-

ence by maintaining a minimum distance between portable and mobile RF communications equipment 

(transmitters) and the Liberator as recommended below, according to the maximum output power of the 

communications equipment.

 

Rated maximum output                         Separation distance according to frequency of transmitter 

 

power of transmitter 

 

  

m

 

150 kHz to 80 MHz 

80 MHz and 800 MHz 

800 MHz to 2,5 GHz

 

 

 d=1.2

d=1.2 

 d=2.3 

P

 

0,01 

0.12 m 

0.12 m 

0.23 m

 

0,1 

0.38 m 

0.38 m 

0.73 m

 

1.2 m 

1.2 m 

2.3 m

 

10 

3.8 m 

3.8 m 

7.3 m

 

100 

12 m 

12 m 

23 m

For transmitters rated at a maximum output power not listed above, the recommended separation dis-

tance (d) in meters (m) can be estimated using the equation applicable to the frequency of the transmit-

ter, where P is the maximum output power rating of the transmitter in watts (W) according to the trans-

mitter manufacturer.

NOTE 1 at 80 MHz and 800 MHz, the separation distance for the higher frequency range applies.

NOTE 2 These guidelines may not apply in all situations. Electromagnetic propagation is affected by 

absorption and reflection from structures, objects and people.

* This table is included as a standard requirement for equipment which has been tested to specific test 

levels and over specific frequency ranges and been found compliant with regulations.

Table 3

Guidance and manufacturers declaration—electromagnetic immunity

The Liberator is intended for use in the electromagnetic environment specified below. The customer or 

the user of the Liberator should assure that it is used in such an environment.

  Immunity test 

IEC 60601 test level 

Compliance level 

Electromagnetic environment—guidance

  Electrostatic  

+- 8kV Contact 

+- 8kV Contact 

Floors should be wood, concrete or ceramic

  discharge (ESD) 

+- 2 kV, +-4 kV,   

+- 2 kV, +-4 kV, 

tile. If floors are synthetic, the relative

  IEC 61000-4-2 

+- 8 kV, +- 15 kV air 

+- 8 kV, +- 15 kV air  humidity should be at least 30%.**

  Electrical fast  

±2 kV for power 

Not applicable 

Not applicable

  transient/burst  

supply lines 

DC powered device 

  IEC 610004-4 

±1 kV for  

Not applicable  

  

input/output lines  No data input/output lines

   Surge 

±1 kV line(s) to line(s) 

Not Applicable 

Not Applicable

   IEC 61000-4-5 

±2 kV line(s) to earth 

DC powered device

Voltage dips,  

<5% UT (>95% dip in UT)  

short interruptions  

for 0,5 cycle 

 

and voltage  

40% UT (60% dip in UT) 

Not Applicable 

Not Applicable  

 

variations on  

for 5 cycles 

DC powered

power supply  

70% UT (30% dip in UT)  

device 

 

input lines  

for 25 cycles 

 

IEC 61000-4-11  <5% UT (>95% dip in UT) for 5 sec

  Power frequency 

30 A/m 

30 A/m 

Power frequency magnetic fields should be

  (50/60 Hz) 

  50/60 Hz   

50/60 Hz 

that of a typical commercial or hospital

  magnetic field 

  

 

environment.

  IEC 61000-4-8 

Note: UT is the a.c. mains voltage prior to application of the test level.

** This statement indicates that the required testing was performed in a controlled environment and the 

Liberator are found to be compliant with regulations.

Summary of Contents for Liberator 20

Page 1: ...Liberator User Manual English...

Page 2: ...an Community If the product unique device identifier UDI label has the CE symbol on it the device complies with the requirements of Directive 93 42 EEC concerning medical devices The CE symbol indicat...

Page 3: ...kg 48 6 lb 22 04 kg 50 lb 22 68 kg 55 lb 24 95 kg 75 4 lb 34 19 kg Weight Filled 89 7 lb 40 69 kg 123 36 lb 56 13 kg 141 01 lb 63 98 kg 165 32 lb 74 99 kg 220 68 lb 100 1 kg Height 24 5 in 622 mm 29 5...

Page 4: ...URRED Caution Use the Liberator only as directed by your doctor Caution The unit contains liquid oxygen which is extremely cold almost 300 F 184 C Exposure to such a low temperature can cause severe f...

Page 5: ...unit which provides you with supplementary oxygen as prescribed by your physi cian This user manual contains the instructions for using the Liberator Refer to the user manual supplied with the portab...

Page 6: ...DOES NOT SEPARATE EASILY DO NOT USE FORCE THE UNITS MAY BE FROZEN TOGETHER LEAVE THE UNITS CONNECTED AND WAIT UNTIL THEY WARM UP THEN THEY WILL SEPARATE EASILY DO NOT TOUCH ANY FROSTED PARTS Caution S...

Page 7: ...Indicator lights up when the button is depressed inform your health care pro vider the next time your Liberator is filled Gen 4 Meter 5 Install the DISS extension 6 Either a Attach cannula to the DISS...

Page 8: ...g and empty as required 4 If the Portable runs low before the Liberator is fully defrosted you may refill it as needed Battery Care and Maintenance Depress button to display level If level is displaye...

Page 9: ...to time to relieve excess pressure especially soon after filling If hissing is persistent or abnormal this could indicate excess pressure being vented off or a leak in the system Contact service prov...

Page 10: ...be stored in the upright position and be well ventilated Do not allow the device to lie on its side Humidity up to 95 noncondensing Temperatures range from 40 F to 158 F 40 C to 70 C Operating temper...

Page 11: ...ovided in this manual Caution Portable and mobile radio frequency RF communications equipment can affect Medical Electrical Equipment Caution The Liberator Reservoir should not be used adjacent to or...

Page 12: ...r equipment which has been tested to specific test levels and over specific frequency ranges and been found compliant with regulations Table 3 Guidance and manufacturers declaration electromagnetic im...

Page 13: ...ned by an electromagnetic site surveya should be less than the compliance level in each frequency rangeb Interference may occur in the vicinity of equip ment marked with the following symbol Test freq...

Page 14: ...14 ENG PN MN234 G User Manual Liberator Notes...

Page 15: ...PN MN234 G User Manual 15 ENG Liberator Notes...

Page 16: ...5535 0989 Email medstar medstar com br CAIRE and CAIRE Inc are registered trademarks of CAIRE Inc Please visit our website below for a full listing of trademarks Trademarks www caireinc com corporate...

Reviews: