background image

Go to www.CAIREmedical.com for more information.

Copyright © 2014 Chart Industries  ML-CONC0057 A

CAIRE Inc. reserves the right to discontinue its products, or change the prices, materials, equipment, quality, 
descriptions, specifications and/or processes to its products at any time without prior notice and with no further 
obligation or consequence. All rights not expressly stated herein are reserved by us, as applicable.

All intake and exhaust locations should remain un-obstructed. Ensure the Companion 5 is located at least 12 in (30.5 cm) away from all walls, 
draperies, furniture, etc.  

Air Intake Filter

The air intake filter is located underneath the Companion 5. It should be replaced once every 2 years as part of preventative maintenance 
or more frequently as needed. It is accessed through a filter door on the bottom of the unit. A single screw must be removed with a Phillips 
screwdriver to unlock the filter door. 

HEPA Filter

The HEPA filter is located inside of the Companion 5 on the front case half of the unit. It should be replaced once every 2 years or more 
frequently as needed.  

Oxygen Concentration 

The Companion 5 will produce oxygen at 90% (+5.5/-3%) at all flow rates. 

CAIRE recommends that the oxygen concentration be checked by the provider every 6 months on units without OCSI (Oxygen Monitoring). 
This is not required on units with OCSI (oxygen monitor). The provider may check the concentration more frequently as required.   

Hour Meter

The LCD hour meter is clearly visible to the right of the flow meter on the front of the Companion 5. It displays the cumulative hours to the 
nearest tenth and cannot be re-set. 

The hour meter will display a diagnostic alarm code if the Companion 5 is in an alarm condition. 

Alarms

Condition

Visual Indicator

Audible Indicator

Alarm Code

None

Green Solid Light

None

None

System Malfunction

Red Solid Light

Continuous Alarm

AL-P01

Power Loss

Red Solid Light

Continuous Alarm

AL-P02

Low Flow

Red Flashing Light

Beeping

AL-P20

High Flow

Red Flashing Light

Beeping

AL-P40

Low O2 Concentration (<85%)**

Yellow Flashing Light

Beeping

AL-P08

Low O2 Concentration (<70%)**

Yellow Solid Light

Continuous

AL-P04

**Only for OCSI units**.

ultraSILENT

 Technology

Companion 5 is designed to produce the quietest air intake and exhaust (purge) cycles.
 

• A patent-pending intake filter is combined with a 2-chamber muffler to reduce the intake sound. 

 

• A purge muffler to quiet the exhaust gas produces an industry-leading quietest purge cycle that virtually eliminates the exhaust noise. 

autoFLOW

 Technology

Companion 5 features a patent-pending variable valve timing system that adjusts the valve cycle timing based on flow rate.
 

• Reduces power by 20% to 250W @ 2.0 LPM

 

• Reduced operating temperatures, system pressures, sound levels, and component strain at lower flow rates. 

Reviews: