Oxygen Concentration 

The VisionAire 3 & 5 will produce oxygen at 90% (+5.5/-3%) at all flow rates. 
The VisionAire 2 will produce oxygen at 95% (+0.5/-3%) up to 1 LPM, and at 90% (+5.5/-3%) between 1 LPM and 2 LPM.
CAIRE recommends testing the oxygen concentration upon delivery to the patient and at periodic intervals set by the equipment provider. The 
provider may check the concentration as frequently as required.
 

Hour Meter

The LCD hour meter is clearly visible under the flow meter on the left of the front panel. 
The hour meter is dual function. The first mode tracks the total cumulative hours. This listing cannot be reset. It also has a second mode to 
calculate accumulated hours. It is designated by “TMR1” on the hour meter and can be reset. 
Toggle between the total cumulative hours and accumulated hours by pressing the pin-hole button on the hour meter. On the TMR1 screen, 
you can re-set the accumulated hours counter by pressing and holding the pin-hole button for 3 seconds.

Alarms

Alarms

 

Condition

Audible Indicator

Visual Indicator

No Power

1 Beep 

-

Low O2 Concentration**

2 Beeps

Yellow Light

Low Pressure

3 Beeps

-

High Pressure

4 Beeps

-

High Temperature

5 Beeps

-

No Flow

6 Beeps

-

 **Only on units with optional Oxygen Monitor**.