Caen V993 Technical Information Manual Download Page 4

  

 

 

 

 

 

 

 

 

Document type: 

Title: 

Revision date: 

Revision: 

User's Manual (MUT) 

Mod. V993 B Dual Timer 

03/03/2004 

 
 
 

NPO: 

Filename: 

Number of pages: 

Page: 

00114/00:V993x.MUTx/03 V993B_REV3.DOC 

10 

 

1. Module  description 

1.1. Overview 

 

The 

Model V993 B

 Dual Timer is a 1-unit VME module housing two identical triggered 

pulse generators. 
The module produces NIM, TTL and ECL pulses whose width ranges from 50 ns to 10 s 
when triggered. Output pulses are provided normal (

OUT

) and negated (

/OUT

). 

Timers can be re-triggered with the end marker signal.  
The coarse adjustment of the output width is provided via a 9-position rotary switch, the 
fine adjustment can be performed via either a rotary handle or by providing an external 
voltage (

WSET

).  

 
The trigger 

START

 can be provided via either an external signal (NIM, TTL  or ECL) or 

manually via a front panel switch (

START/RESET

 switch). 

The 

VETO

 input (NIM, TTL  or ECL TRUE level) allows to disable the START operation. 

The 

ENDM

 output (NIM, TTL  or ECL) is a short pulse produced at the end of any timing 

cycle. 
The 

RESET

 input (NIM, TTL  or ECL) ends the timing cycle at any time. RESET can also 

be provided manually via a front panel switch (

START/RESET

 switch). 

See § 2.4 for details about signals. 
 
The module features LEMO 00 connectors for WSET and NIM/TTL signals and male pin 
couples for ECL signals. Two front panel LEDs display when the module is operating. 
The board is equipped with the P1 connector. 
 
Two internal switches allow to select between NIM and TTL output pulses, while input 
levels are automatically recognised (no selection is required). 
 

 

Summary of Contents for V993

Page 1: ...Technical Information Manual MOD V993 3 March 2004 Revision n 3 DUAL TIMER MANUAL REV 3 NPO 00114 00 V993x MUTx 02...

Page 2: ...he User accident or any abnormal conditions or operations CAEN declines all responsibility for damages or injuries caused by an improper use of the Modules due to negligence on behalf of the User It i...

Page 3: ...PECIFICATIONS 5 2 1 PACKAGING 5 2 2 POWER REQUIREMENTS 5 2 3 FRONT PANEL 6 2 4 SIGNALS FEATURES 7 2 5 OTHER COMPONENTS 8 2 6 TECHNICAL SPECIFICATIONS TABLE 8 3 OPERATING MODES 9 3 1 INSTALLATION 9 3 2...

Page 4: ...roviding an external voltage WSET The trigger START can be provided via either an external signal NIM TTL or ECL or manually via a front panel switch START RESET switch The VETO input NIM TTL or ECL T...

Page 5: ...V993x MUTx 03 V993B_REV3 DOC 10 5 2 Technical specifications 2 1 Packaging The Model V993 B is housed in a 6U high 1U wide VME unit The board hosts the VME P1 connector 2 2 Power requirements The powe...

Page 6: ...Model V993 B front panel DUAL TIMER ST VETO RESET ENDM OUT OUT ST VETO RESET ENDM OUT OUT RESET RESET START OUT OUT ENDM OUT OUT ENDM START VETO START VETO WSET WSET 1 s 10 s 1 ms 1 ms 10 ms 10 s 1 s...

Page 7: ...sensitive 3 ns minimum FWHM on LEMO 00 type connector ECL Leading edge sensitive 3 ns minimum FWHM on a male pin couple START RESET Switch2 Momentary switch Function It stops the timing cycle at any t...

Page 8: ...age to the input stage 2 5 Other components OPERATION LEDS 2 green red LEDs see 3 5 WSET FINE ADJUSTMENT Two 2 position switches see 3 3 COARSE ADJUSTMENT Two 9 position rotary switches see 3 3 FINE A...

Page 9: ...NEUTRALIZE ANY STATIC ELECTRIC CHARGE BUILT UP ON THE BODY e g TOUCHING AN EARTHED OBJECT BEFORE HANDLING THE ECL CONNECTORS CAUTION ALL CABLES MUST BE REMOVED FROM THE FRONT PANEL BEFORE EXTRACTING T...

Page 10: ...he WSET connector with a DC voltage ranging from 0 to 5 Volts the WSET FINE ADJUSTMENT SWITCH allows to select between the two ways Left WSET Right Rotary Handle see 2 5 3 4 Output level selection Two...