A1539, AG539, A154X, AG54X  Power Supply Boards 

3.

 

Safety and installation requirements 

General safety information 

This section contains the fundamental safety rules for the installation and operation of the board. Read 
thoroughly  this section before starting any procedure of installation or operation of the product. 

Injury Precautions 

Review the following precautions to avoid injury and prevent  damage to this product or any products 
connected to it. To avoid potential hazards, use the product only as specified. Only qualified personnel 
should perform service procedures. 

Avoid Electric Overload. 

To avoid electric shock or fire hazard, do not power a load outside of its specified range. 

Avoid Electric Shock.

 

To avoid injury or loss of life, do not connect or disconnect cables while they are connected to a voltage 
source. 

Do Not Operate Without Covers.

 

To avoid electric shock or fire hazard, do not operate this product with covers or panels removed. 

Do Not Operate in Wet/Damp Conditions. 

To avoid electric shock, do not operate this product in wet or damp conditions. 

Do Not Operate in an Explosive Atmosphere. 

To avoid injury or fire hazard, do not operate this product in an explosive atmosphere.

 

Do Not Operate With Suspected Failures.

 

If you suspect this product to be damaged, have it inspected by qualified service personnel. 

Safety Terms and Symbols on the Product 

These terms may appear on the product: 
DANGER indicates an injury hazard immediately accessible as you read the marking. 
WARNING indicates an injury hazard not immediately accessible as you read the marking. 
CAUTION indicates a hazard to property including the product. 
The following symbols may appear on the product: 

 

DANGER 

 

WARNING 

High Voltage 

Refer to Manual 

Installation 

The Mod. A1539, AG539, A154X, AG54X are SYx527 boards. At power ON the SYSTEM, the processor will 
scan all the slots in the crate to find out where the module is plugged and what kind of module it is.  

Summary of Contents for A1539

Page 1: ...CAEN Tools for Discovery n Electronic Instrumentation Rev 0 Es A1539 AG539 A154X AG54X Power Supply Boards Rev 2 6 May 2015...

Page 2: ...ccident or any abnormal conditions or operations CAEN declines all responsibility for damages or injuries caused by an improper use of the Modules due to negligence on behalf of the User It is strongl...

Page 3: ...safety information 9 Injury Precautions 9 Safety Terms and Symbols on the Product 9 Installation 9 4 Operating modes 10 Output control and monitoring 10 Output Enable 11 Grounding specifications 12 Sa...

Page 4: ...rs 24 channel SHV versions are double width boards 10 TE all other versions are single width 5 TE All boards are provided with both current and voltage protections If over current occurs the relevant...

Page 5: ...nels 0 500 V common to all board channels VMAX hardware accuracy 1 V VMAX software 0 100 V settable for each channel 0 500 V settable for each channel VMAX software resolution 1 V Ramp Up Down 1 50 Vo...

Page 6: ...single width 5 TE height is 6U External connections The function and electro mechanical specifications of the external connectors are listed in the following subsections Version 24 32 channel DB25 12...

Page 7: ...ntation A1539 AG539 A154X AG54X Power Supply Boards 7 DB25 connector pin assignment Fig 2 DB25 connector pin assignment 32 channel version INT INT interlock RTN channel return AGND earth SH connector...

Page 8: ...s VMAX trimmer Function it allows to adjust the hardware maximum voltage VMAX common to all the channels Its value can be read out via software Shield CFG jumpers 24 32 ch floating Function see table...

Page 9: ...s To avoid electric shock or fire hazard do not operate this product with covers or panels removed Do Not Operate in Wet Damp Conditions To avoid electric shock do not operate this product in wet or d...

Page 10: ...ameter shows the ON OFF channel status As this parameter is set ON the channel is switched on if the INTERLOCK is not active and if the channel is enabled either locally or remotely highlighted in gre...

Page 11: ...r it is possible to monitor board parameters such as measured Temperature and HVMax and to check board status the following messages may be returned by the POWER SUPPLY SYSTEM when monitoring the boar...

Page 12: ...is available as front panel 2mm socket connector This feature allows on detector grounding thus avoiding loops which may increase noise level FAGND and AGND may be coupled in several ways according to...

Page 13: ...HV CABLE AGND 2 1 4 3 6 5 FRONT PANEL JUMPERS 1 HV Generator Load PMT 0 HV Channel 0 RETURN Load PMT 31 HV Channel 31 EARTH of DETECTOR Cable SHIELD Circuit Protected SHIELD Circuit Protected RETURN...

Page 14: ...including engineers scientists and technical professionals who all trust them to help achieve their goals faster and more effectively CAEN S p A CAEN GmbH CAEN Technologies Inc Via Vetraia 11 Klingen...

Reviews: