Disinfecting Scanner and Replacing Disposable Sleeve
To avoid cross contamination between patients, the scanner must be disinfected and the
disposable plastic sleeve must be replaced.
First use disinfectant wipes to clean the scanner arm, cradle, keyboard and mouse. Then
proceed with the steps below in order to replace the disposable sleeve.
STEP 1
Press firmly on both sides of the disposable
sleeve to release.
STEP 2
Pull the disposable sleeve slowly off the scanner
arm and discard.
STEP 3
Note that as the sleeve is removed, the optical
lens is automatically covered over by a small
metal plate. At this time you may slide on a new
sleeve for the next patient.
WARNING !!
When removing the sleeve, be careful not to
touch the optical lens directly with your fingers.
Touching it may cause damage to the optical
lens.