C2Dx STIC Instruction Manual Download Page 2

Follow sterile procedures.

NOTE: ALWAYS start procedure with a full syringe.

1.

 

Perform steps 1 through 6 of the Quick Pressure Moniter Set 
instructions. 

2.

 

Remove the needle from the monitor. 

3.

 

Attached the slit catheter to the tapered chamber stem. Make 
sure the fit is very snug.

NOTE:  
Optional extension  
tubing may be used  
between the slit  
catheter and the  
monitor if  
additional length  
is desired.

 

4.

  Remove the need guard from the break-away needle, and push 

the catheter in until it is just visible at the end of the needle.

CAUTION: DO NOT let the catheter protude from the needle, as it 
may be cut by the needle upon insertion.

 

5.

 

Hold the needle approximately 45° up from horizontal. Slowly 
force the fluid through the disposable to remove air. 

6.

  Bring the catheter tip (slit end) to the same height as the 

disposable diaphragm. Display should read -8 to +9mm/Hg.

 

 

 

 

7.

 

Press the “ZERO” button. 

NOTE: Display must read “00” before continuing. If it does not, 
please see troubleshooting.

8.

 

Hold the catheter and the needle together and push into the 
compartment. Make sure the catheter’s tip is protected inside 
the needle.

9.

 

Push the catheter foward while pulling back on the needle. 
Tape the catheter in place after the needle is out of the skin. 

 

 

Remove the break-away needle. 

NOTE: Pinch the green fins on the needle together, then pul them 
apart and split the needle away.

 

11.

  Keep the chamber in the monitor at the same height as the 

estimated height of the slit tip of the catheter in the body. 
See below for suggested set-up for the Pressure Monitor with 
Indwelling Slit Catheter Set.

 

 

 

 

12.

  Inject less than 3/10 cc of saline into the compartment to 

equillibrate with interstitial fluids.

13

.

  Wait for the display to reach equillibrium.

14

.

  Record the pressure readings at the appropriate intervals. 

Repeat step 12 to ensure patency for subsequent 
measurements.

WARNING: The Indwelling Slit Catheter 
procedure must not exceed a 24 hour duration.

WARNING: Device must be at correct angle of 
insertion before the “ZERO” button is pressed 
to insure accuracy.

WARNING: The monitor must be securely taped 
to the patient to prevent its displacement and/
or patient injury.

WARNING: DO NOT use pressure readings when 
display indicates low batter. Replace battery 
and repeat steps to obtain pressure reading.

← 

Acute 

and Sports 
Medicine  
Set-up 

[Optional 
extentsion 
tubing not 
shown but may 
be necessary.] 
 
Patient shown 
standing.

←  

Alternate 
Acute  
Set-up Only 

[Cannot be 
used for a 
standing 
patient.] 
 
Patient shown 
laying down.

INDWELLING SLIT CATHETER SET INSTRUCTIONS