ENGLISH

micro-gloss  

Operation

mode wheel:

  Switch on 

press

  Show menu 

press

  Select function 

scroll

  Activate function 

press

operate-button:

  Measure

operate

How to calibrate?

Gloss

  Turn on the unit in holder

  Autodiagnosis checks standard  

  and calibration

  If requested, clean the standard  

  and start calibration by pressing  
  the mode wheel

*It is advisable, to perform the zero calibration on the uncoated  
  metal of the sample to be tested.

 Main menu 

 

  Mode 

 

Geometry/Sensor 

}

 

  Memory 

}

 

  Difference 

}

 

  Calibration 

}

 

  Setup 

 *

}

¢

Navigation Symbols:

  Back to previous menu

}

  Got to sub-menu

3

  Function activated

*

  Guide to language menu

¢

Thickness 

(only micro-TRI-gloss μ)

Autodiagnosis

Please clean standard

Confirm 

¢

 [mode]

Cancel 

¢

 [operate]

Main menu
  Calibration                                

}

    Calibrate                                 

}

      Thickness Fe                          

}

      Thickness NFe                       

}

  Select the desired sensor 

  (Fe or NFe)

  Put the unit on the zero  

  standard or original substrate*

  Press operate button

  Hold the unit free in air and  

  press operate