28
Appendix B
CUSTOM SPINDLES
The CAP Series Viscometer allows for the use of custom spindles. Please contact
BYK-Gardner’s Technical Sales Department for more information.
Custom spindles are calibrated using spindle entry codes 11-20. The calibration of a
custom spindle requires the entry of two parameters:
SMC
and
SRC
. The SMC and
SRC are required for proper calculation of viscosity.
The SMC and SRC must be entered after entering the spindle number (11-20) and
selecting
YES (1)
for calibration. Follow the screen prompts.
Once SMC and SRC are entered, follow the procedure described above for spindles
01-10.
NOTE:
SMC and SRC need to be entered only once for any spindle number 11-20.
The CAP Series Viscometer will store these values. SMC and SRC can be
overwritten by executing a spindle calibration.
CONE (00) CALIBRATION
VISCOSITY
FSR
TEMP
S.STRESS
S.RATE
SPEED
TIMER
CONE
SAMPLE
(POISE)
(%)
(Deg C)
(D/CM2)
(1/sec)
(RPM)
(SEC)
No.
No.
Operator:
Date:
Model/Serial#:
Fluid:
Figure B-1