5. Calibration
State-of-the-art technology and the
reference-beam principle mean that
calibration is needed only
infrequently. We recommend a
calibration interval of two months.
Calibration is also necessary:
• at the first startup
• in the event of a drastic change in
ambient temperature (for example,
after moving the unit to a new site)
• if "Please Recalibrate" appears on
the display (for example, after
stored calibration values have
been modified).
To ensure accurate calibration, use
only original standards from
BYK-Gardner.
Do not touch the reference surface.
Protect it against scratching.
Even with careful treatment,
ambient conditions may alter a
standard’s values over time. For
that reason, have Customer Service
check your standards regularly
(once a year) against primary
standards.
Make sure the values displayed
during calibration match the data for
the standards you use, in order to
avoid mixing up or accidentally
changing the calibration values (see
Section 5.3).
To calibrate:
CAUTION: Make sure the
calibration standards are clean
and free from scratches!
5. Calibration
State-of-the-art technology and the
reference-beam principle mean that
calibration is needed only
infrequently. We recommend a
calibration interval of two months.
Calibration is also necessary:
• at the first startup
• in the event of a drastic change in
ambient temperature (for example,
after moving the unit to a new site)
• if "Please Recalibrate" appears on
the display (for example, after
stored calibration values have
been modified).
To ensure accurate calibration, use
only original standards from
BYK-Gardner.
Do not touch the reference surface.
Protect it against scratching.
Even with careful treatment,
ambient conditions may alter a
standard’s values over time. For
that reason, have Customer Service
check your standards regularly
(once a year) against primary
standards.
Make sure the values displayed
during calibration match the data for
the standards you use, in order to
avoid mixing up or accidentally
changing the calibration values (see
Section 5.3).
To calibrate:
CAUTION: Make sure the
calibration standards are clean
and free from scratches!